Device Maintenance
Balloon Maintenance:
Balloon Fill Volume Table:
20
Routine maintenance is recommended for the AMT G-JET
®
Button.
Check the balloon ll volume as recommended by your healthcare provider. While holding the
external bolster (see page 6, #1) in place, insert a syringe into the balloon ll valve (see page 6,
#2) and withdraw the water to examine the amount. Rell the balloon with the originally prescribed
amount of water or as prescribed by your healthcare provider.
French Size
14FR
16FR
18FR
Minimum
3ml
3ml
7ml
Recommended
4ml
4ml
8ml
Maximum
5ml
5ml
9ml
Adjustable Balloon:
People do not come in one size, so neither should their balloons! Your healthcare provider can
recommend the best ll volume for you. Just like AMT’s MiniONE
®
Balloon Buttons, the G-JET
®
Button’s balloon ll volume can be easily adjusted for an individualized-t to reduce leaks and
improve patient comfort. See below for minimum, recommended, and maximum ll volumes.
5 ml4 ml3 ml
Applied Medical Technology, Inc.
Feed Set & Extension Set Maintenance
21
Feed Sets and Extension Sets should be cleansed with mild soap and warm water
after each use.
Be sure to rinse thoroughly to remove any formula or soap residual.
For continuous feeding, the Feed Set should be cleaned at least once per day.
To prevent clogging, always ush the Feed Set & Extension Set after use.