What to do if Anilam RBS-T DRO shows no movement?
- NNicholas MartinezAug 2, 2025
If your Anilam Media Converter DRO shows no movement, and the display fluctuates between 0 and the resolution setting, it could be due to several reasons. The reader head might be over torqued or improperly aligned, or the reader head alignment brackets may not have been removed. A damaged cable could also be the cause. To resolve this, verify the encoder and reader head are secure, reinstall the plastic alignment brackets and remount if necessary. You can also try swapping cable connections at the DRO to see if the problem moves to another axis. If the issue persists, consider replacing the reader head.