Lovis 2000 M/ME Low Volume – SOP
Standard Operating Procedure for Low Sample Volume
1 About the SOP
Read all safety instructions in the Lovis 2000 M/ME Instruction Manual (Original Instruction)!
Find additional and more detailed information in the Lovis 2000 M/ME Instruction and Reference Manual
and in the Instrument Firmware Manual.
2 Conditioning the Short Capillaries and Plugs
Condition every capillary before using it for the first time. Clean every new hollow plug to remove
particles that would interfere with the measurement.
TIP: Select a suitable cleaning liquid. It must completely dissolve your sample, evaporate easily and
without residues. It must not react chemically with the capillary, the plugs or any of the wetted
parts. Consult the Lovis
2000 M/ME Instruction Manual for more information.
5. Blow the capillary dry with the built-in air pump:
Plug a clean hollow needle 1.2 (gauge 18) onto the Adapter Luer cone (13) of the air hose.
2 .... Capillary
6 .... Hollow needle
13.... Adapter Luer cone of air hose
Insert the hollow needle into the
capillary as far as possible.
Start the air pump with the button
on the touch screen.
The drying time depends on the cleaning liquid, on the temperature and on the ambient air
humidity. Determine the suitable drying time empirically.
Stop the pump with the button .
This is a step-by-step guide how to... fill a short capillary with a not too highly viscous liquid
(max. 200
mPa.s) and a steel ball
perform a single-point measurement (at one angle and
one temperature) and
clean the short capillary.
1. Fill some cleaning liquid into the short capillary (2)
with a syringe (5) with a hollow needle (6).
2. Push a supplied brush (12) as far as possible into
the short capillary and rotate it. Move it up and
down repeatedly.
3. Rinse the capillary with some more cleaning liquid
as described above.
4. Rinse the brush and wipe it with a lint-free cloth or
tissue paper.
2 .... Capillary
5 .... Syringe
6 .... Hollow needle
12 .... Brush
6
5
2
12
6
13
26
Question and Answer IconNeed help?

Do you have a question about the Anton Paar Lovis 2000 M/ME and is the answer not in the manual?

Anton Paar Lovis 2000 M/ME Specifications

General IconGeneral
BrandAnton Paar
ModelLovis 2000 M/ME
CategoryLaboratory Equipment
LanguageEnglish

Summary

About the SOP

Conditioning the Short Capillaries and Plugs

Installing Lovis Measuring Systems

Filling the Short Capillaries

Performing a Level Adjustment

Adjusting Lovis Measuring Systems

Creating and Editing Measuring Methods

Measure

Cleaning

Cleaning the ball

Procedure for cleaning the measuring ball, emphasizing careful handling to prevent contamination.

Cleaning the microliter glass syringe and hollow needles

Instructions for cleaning the microliter glass syringe and hollow needles to avoid sample contamination.

Brushing short capillaries

Steps for brushing short capillaries using a supplied brush and cleaning liquid.