Guidance and manufacturer's declaration
- Electromagnetic immunity
The Arm Blood Pressure Monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of Arm Blood
Pressure Monitor should assure that it is used in such an environment.
Immunity test
test level
Compliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
±2 kV,±4kV,±8 kV,
±15kV air
±8 kV contact
±2 kV,±4 kV,
±8 kV,±15 kV air
Electrostatic
discharge (ESD)
IEC 61000-4-2
±2 kV,±4 kV, ±8 kV,±15
kV air
±8 kV contact
±2 kV,±4 kV,±8 kV,
±15 kV air
Electrical fast
transient/burst
IEC 61000-4-4
Not applicable
Surge
IEC 61000-4-5
Not applicable
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines IEC
61000-4-11
Not applicable
Power frequency
Magnetic field
IEC 61000-4-8
30A/m, 50/60Hz
Conducted RF
IEC61000-4-6
Not applicable
Radiated RF
IEC61000-4-3
80 MHz - 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz - 2,7 GHz
80 % AM at 1 kHz
NOTE: UT is the a.c. mains voltage prior to application of the test level