Do you have a question about the Apex Digital DIGI-STIM and is the answer not in the manual?
| Brand | Apex Digital |
|---|---|
| Model | DIGI-STIM |
| Category | Fitness Equipment |
| Language | English |
This device is a prescription device and only for symptomatic relief of chronic intractable pain.
Lists situations or conditions where the device should not be used, such as with pacemakers or on the carotid sinus.
Covers general safety advice, including keeping the device away from children and avoiding use during pregnancy.
Details potential risks like skin irritation, interference with medical equipment, and advice on electrode placement.
A protective cover that conceals controls for Mode, Time, Width, and Rate adjustments.
Adjusts the strength of stimulation and functions as ON/OFF controls, located under the cap.
Used to select treatment types like Burst, Continuous, Modulation, SDR, and SDW.
Allows pre-selection of treatment duration: 15, 30, 60 minutes, or continuous output.
Regulates the pulse width for both channels of the device.
Controls the number of pulses per second for both channels.
Guides on how to adjust intensity knobs for Channel 1 and Channel 2 for comfortable stimulation.
Specifies the waveform used by the device as Asymmetrical Bi-phasic square pulse.
Indicates the voltage range of the device is 0-110 Volt (Open Circuit).
Specifies the maximum charge per pulse as 21 micro-coulombs.
States the device is powered by a 9-Volt Battery.
Provides the physical dimensions of the device: 108mm(H) ×61.5mm(W) ×25mm(T).
Indicates the weight of the device, including the battery, is 140 grams.
Cycle Bursts mode with 2 Bursts/sec, 9 pulses/Burst, 100Hz, and adjustable width.
Continuous mode with adjustable pulse rate, pulse width, and intensity.
Modulated Width mode where pulse width varies automatically over 6 seconds.
Strength-Duration and Rate modulation mode with alternating intensity and pulse rate.
Strength-Duration and Width modulation mode with alternating intensity and pulse width.
Guidance on maintaining minimum distances between the device and RF communications equipment to prevent interference.
Details the device's immunity to electromagnetic fields, including conducted and radiated RF.
Specifies the device's immunity to electrostatic discharge (ESD) based on contact and air discharge levels.
Instructions for physicians on accessing and using special doctor mode features.
Procedure for locking treatment parameters to prevent patient alteration of settings.
Procedure for unlocking previously locked treatment parameters to allow adjustments.