EasyManua.ls Logo

Aquafine VL200 Series - User Manual

Default Icon
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
425 Clair Rd. W, Guelph, Ontario, Canada N1L 1R1
t. (+1) 800.423.3015
e-mail: techsupport@aquafineuv.com
www.aquafineuv.com
Models:
VL200, VL410 Series
Owner’s Manual
520513_RevF

Questions and Answers

  • K
    Kelly MccartyAug 20, 2025
    What to do if the Aquafine VL200 Water System produces heated product water?
    • A
      ariel06Aug 20, 2025
      If the product water from your Aquafine Water System is heated, run the water until it returns to ambient temperature.
  • D
    Derrick WebbNov 4, 2025
    Why is there a pressure drop in my Aquafine Water System?
    • A
      Amy LittleNov 4, 2025
      If you're experiencing a pressure drop in your Aquafine Water System, several factors could be responsible. It might be due to a clogged sediment pre-filter, in which case you should replace the filter cartridge. Also, check your source water supply, as fluctuations in source pressure can occur. Another reason may be a change in feed water quality, so you should have the source water tested to ensure it's within the system's allowable limits. It's also possible there's contamination in the water lines after the UV system. In this case, it is imperative that the effluent water stream be shocked with chlorine (bleach) before water leaves the UV system, because the UV system must have a bacterial free distribution system to work effectively. Finally, there could be a breakthrough of sediment ...

Summary

Section 1 Safety Information

Potential Hazards Overview

Explains various hazard symbols and their meanings.

Essential Safety Precautions

Details critical safety precautions and potential dangers for operation.

Warnings and Cautions

Covers warnings on hot water, UV exposure, modifications, and material cautions.

Water Chemistry Recommendations

Outlines recommended water quality parameters for optimal UV system performance.

Section 2 General Information

System Components and Parts

Lists and describes the main components of the UV system.

Section 3 Installation

UV System Mounting and Placement

Covers mounting the UV system vertically or horizontally.

Installation Procedures and Checks

Details steps for selecting location, mounting, connecting, and leak checks.

System Startup and Cleaning

Explains initial startup cleaning and distribution system disinfection.

Section 4 Maintenance

Replacing the UV Lamp

Step-by-step procedure for replacing the UV lamp after 9000 hours of operation.

Cleaning and Replacing Quartz Sleeve

Detailed procedure for cleaning and replacing the quartz sleeve.

Cleaning and Replacing UV Sensor

Step-by-step guide for cleaning and replacing the UV sensor.

Section 5 Operation

210153 Controller Operation Guide

Explains lamp life, A3 code, reset, total days, and lamp failure for 210153 controllers.

210154-M Controller Operation Guide

Covers UV intensity monitoring, levels, and alarms for 210154-M controllers.

System Output Options

Details auxiliary, solenoid, and optional 4-20mA output functionalities.

Section 6 Troubleshooting

Common Troubleshooting Issues

Addresses common symptoms like pressure drop, bacteria, leaks, and lamp failure.

Maintenance Schedule and Fault Modes

Provides a maintenance schedule and explains display fault modes.

Section 7 Manufacturers Dose Flow Chart

Dose Flow Curves Explained

Presents dose curves for VL410 and VL200 models based on flow rate.

Section 8 Specifications

Performance and TOC Specifications

Details performance and TOC specifications for different models.

Section 9 Manufacturers Warranty

Warranty Information and Contacts

Lists customer service contacts and outlines the equipment warranty terms.

Aquafine VL200 Series Specifications

General IconGeneral
DimensionsVaries by model
WeightVaries by model
ApplicationWater disinfection
Lamp TypeLow-pressure high-output UV lamps
MaterialStainless steel