Model 1092/1093 Program ROM Update
1 Introduction
This document provides instruction to replace the program ROM (firmware) in the Model 1092A/B/C
and Model 1093A/B/C series GPS Satellite Controlled Clocks. Basic understanding of electrical safety and
static protection is expected.
2 Required Equipment
1. Torx T-25 or large flat-blade screw driver
2. IC Removal Tool, curved tweezers or small blade screwdriver
3 Saving Original Configuration
Updating the firmware will erase the clock’s configuration. Document the existing settings for reference.
1. 1092B and 1093B/C: press SETUP and sequentially scroll through the various menus.
2. Press ENTER on any menu to sequence through that menu’s settings.
3. 1092A and 1093A: unfortunately, many settings cannot be queried. Use the Operation Manual for a
list of serial commands to set local offset, Daylight Saving Time, broadcast modes, IRIG-B time and
control functions, etc.
4 Disassembly
1. Disconnect power.
2. Remove from service.
3. Remove the four screws securing the cover with a T-25 driver (or flat-blade screwdriver).
4. Lift off the cover.
5 Installing the Program ROM
1. The program ROM is located at U2 (see Figure 1 on page 4).
2. Remove the old program ROM using the appropriate IC removal tool.
3. Install the new ROM. Observe the correct IC orientation and make sure the IC pins line up well on
both sides of the socket. If necessary, use a pin-bending tool to line up the pins.
4. Option 28: If present, the Power System Time, Frequency, and Phase Monitor may also require an
update ROM. Follow the separate instructions for the update steps.
6 Reassembly
1. Replace the cover.
2. Install the four screws securing the cover with a T-25 driver (or flat-blade screwdriver).
3. Place into service.
4. Connect the power.
5. Set to factory defaults.
7 Configuration
7.1 Factory Defaults
7.1.1 Front Panel (B and C Models)
1. Press and release SETUP at startup and during the first two splash screens.
If not displayed then start over.
2. Display will show SET FACTORY DEFAULTS?
3. Press ENTER and follow the on-screen prompts.
Arbiter Systems, Inc. · 1324 Vendels Circle, Suite 121 · Paso Robles, CA 93446 · USA · Tel: +1.805.237.3831 PD0030600I
Fax: +1.805.238.5717 · E-mail: techsupport@arbiter.com · Internet: http://www.arbiter.com 1

Other manuals for Arbiter Systems 1093B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arbiter Systems 1093B and is the answer not in the manual?

Arbiter Systems 1093B Specifications

General IconGeneral
BrandArbiter Systems
Model1093B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals