III - VH TRIP
®
Calibration :
Your VH TRIP
®
is easy to calibrate.
Step 1 : Probe Calibration.
1. Reset both displays to Zero by pressing reset buttons ( 5 ).
2. Slide Probe calibration switch to the right ( 2 ).
LED ( 3 ) comes on.
3. Turn the wheel on which your probe is installed 3 times (no more – no less).
4. Slide Probe calibration switch back to initial position ( 2 )
5. Switch OFF your VH TRIP
®.
Your VH TRIP
®
is now ready to be calibrated.
If LED ( 3 ) flashes when sliding Probe Calibration Switch back to the left, your
calibration is incorrect. Please refer to the Troubleshooting guide at the end of this
instruction manual.
WARNING: If the Probe Calibration Switch is moved by mistake, the probe calibration
process will have to be repeated.