SL1 DATA LOGGER User Manual Rev. AD
sales.get@athena.euwww.getdata.it
ENGLISH
6.3 POLYLINE CALIBRATION
In the Polyline calibration, a table must be compiled with the various calibration points
Pressing any box in the Calibration Values table makes it possible to change its value.
The Calibration Values table must be filled in according to the following method:
Each value in the Raw Value column must have the corresponding value in the Cal. Value column
where:
The Raw Value is the raw value read from the acquired channel
The Cal. Value is the desired output value corresponding to the Raw Value input value
For example, if the sensor in the acquired channel is an NTC, the Raw Value will be the output
voltage of the NTC whereas the Cal. Value will be the temperature related to that voltage value.