42 APx585/586 B Series Audio Analyzer: Analog I/O Specifications
585
DC Voltage Measurement
Valid only for input bandwidths 90k
Input Ranges 0.32 V to 100 V, 10 dB steps ±120 Vdc maximum in 100 V range
Accuracy
0.32 V range ±(0.7% reading + 800 µV)
1 V–100 V ranges ±(0.7% reading + 0.1% range)
Normal Mode Rejection Typically > 90 dB, 20 Hz to 20 kHz.
NOTES to SPECIFICATIONS:
1 System specification including contributions from both generator and analyzer. Generator-only and/or analyzer-only
contributions are typically less.
2 Generator load must be 600 balanced or 300 unbalanced for specified performance. Generator dc offset
must be off or set to 10 mV.
3 Input must be 150 mV for specified performance. Analyzer must be set to measure “d2+d3” for MOD and DFD.
4 Valid for the balanced input configuration only.
5 Maximum low-pass filter frequency is limited by analyzer input bandwidth setting.
6 Filter response is relative to “no filter” selection; overall system performance will also include analog flatness imper-
fections. DSP warping may significantly increase roll-off rate and lower ENBW.