EasyManua.ls Logo

Avantes AVASPEC-128 - Page 25

Avantes AVASPEC-128
129 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24 Apr-09
Avantes www.avantes.com info@avantes.com
2.4 AvaSpec-2048 Fiber Optic Spectrometer
AvaSpec-2048
The AvaSpec-2048 Fiber Optic Spectrometer is
based on the AvaBench-75 symmetrical Czerny-
Turner design with 2048 pixel CCD Detector
Array. The spectrometer has a fiber optic
entrance connector (standard SMA, others
possible), collimating and focusing mirror and
diffractional grating. A choice of 16 different
gratings with different dispersion and blaze
angles enable applications in the 200-1100nm
range. The AvaSpec-2048 can be delivered with
2 platforms of electronics; either with USB1.1
with 14 bit AD converter or the new USB2.0
interface with 16 bit AD converter.The
AvaSpec-2048 is especially suitable for low light
level and high resolution applications. An optional detector coating enhances the CCD performance
for the UV range and a detector collection lens offers high sensitivity. Digital IO ports enable
external triggering and control of shutter and pulsed light sources from the Avantes line of
instruments.
The AvaSpec-2048 is also available as dual channel or multiple channel instrument (up to 8
channels), where all spectra are taken simultaneously.
The AvaSpec-2048 comes with AvaSoft-basic, a complete manual, USB interface cable and a PS-
12V/1.0A power supply. AvaSoft-full and application software can be ordered separately.
Alternatively the AvaSpec-2048-SPU has a switch to run on USB power or external power.
The new AvaSpec2048-USB2 has a USB2 interface with ultrafast datasampling of 500 spectra per
second and datatransfer in 2msec and supports analog in-and outputs as well. Optional Bluetooth
(-BT) communication and an SD card for on-board saving of spectra can be added. The AvaSpec-
2048-USB2 runs on USB power and comes with AvaSoft-basic, a complete manual and USB interface
cable. Multiple (up to 127) USB2 spectrometers with different detector types can be externally
coupled.
Technical Data
Spectrometer platform
AvaSpec-2048
AvaSpec-2048-USB2
Optical Bench
Symmetrical Czerny-Turner, 75 mm focal length
Wavelength range
200-1100 nm
Resolution
0.04 20 nm, depending on configuration (see table)
Stray light
< 0.1%
Sensitivity (AvaLight-HAL, 8 µm fiber)
counts/µW per ms integration time
5000 (14-bit AD converter)
20,000 (16-bit AD converter)
Detector
CCD linear array, 2048 pixels
Signal/Noise
200:1

Related product manuals