EasyManua.ls Logo

Avantes AVASPEC-128 - Page 31

Avantes AVASPEC-128
129 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30 Apr-09
Avantes www.avantes.com info@avantes.com
Grating selection table for AvaSpec-3648
Use
Useable range
Spectral range
(nm)
Lines/mm
Blaze (nm)
Order code
UV/VIS/NIR
200-1100**
900**
300
300
UA
UV/VIS
200-850
520
600
300
UB
UV
200-750
250-220*
1200
250
UC
UV
200-650
165-145*
1800
UV
UD
UV
200-580
115-70*
2400
UV
UE
UV
220-400
70-45*
3600
UV
UF
UV/VIS
250-850
520
600
400
BB
VIS/NIR
300-1100**
800**
300
500
VA
VIS
360-1000
500
600
500
VB
VIS
300-800
250-200*
1200
500
VC
VIS
350-750
145-100*
1800
500
VD
VIS
350-640
75-50*
2400
VIS
VE
NIR
500-1050
500
600
750
NB
NIR
500-1050
220-150*
1200
750
NC
NIR
600-1100**
500**
300
1000
IA
NIR
600-1100
500
600
1000
IB
* depends on the starting wavelength of the grating; the higher the wavelength, the bigger the dispersion and the smaller
the range to select.
** please note that not all 3648 pixels will be used for the useable range
Resolution table (FWHM) for AvaSpec-3648
*depends on the starting wavelength of the grating; the higher the wavelength, the bigger the dispersion and the better the resolution
Slit size (µm)
Grating
(lines/mm)
10
25
50
100
200
500
300
0.5
1.4
2.4
4.3
8.0
20.0
600
0.27
0.7
1.2
2.1
4.1
10.0
1200
0.07-0.13*
0.2-0.3*
0.4-0.6*
0.7-1.0*
1.4-2.0*
3.3-4.8*
1800
0.05-0.08*
0.12-0.21*
0.2-0.36*
0.4-0.7*
0.7-1.4*
1.7-3.3*
2400
0.04-0.07*
0.08-0.15*
0.14-0.25*
0.3-0.5*
0.5-0.9*
1.2-2.2*
3600
0.025-0.04*
0.07-0.10*
0.11-0.16*
0.2-0.3*
0.4-0.6*
0.9-1.4*

Related product manuals