EasyManua.ls Logo

Avantes AVASPEC-128 - Page 41

Avantes AVASPEC-128
129 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40 Apr-09
Avantes www.avantes.com info@avantes.com
2.10.2 USB2 platform connections
Power LED green and scan LED yellow
The green and yellow LED´s act as status LED´s for the micro controller with following meaning:
Green LED = off, power is not connected
Green LED = on, power is on, micro controller ready, no errors
Green LED = blinking, permanent error detected by micro controller
Yellow LED = on, when scan is transmitted to PC
Power connector (only needed for RS-232 functionality with SPU2)
The power connector is a Low power DC connector with GND on outer contact and +12V on inner
contact. The outside diameter is 5.5mm, the inside diameter 2.1mm.
The electrical circuit accepts voltages between 5 and 15V.
NOTE: Please use Avantes PS-12VDC/1.0A power supply or 12VDC batterypack only, serious damage
to the electronics may occur, when other power supplies with different polarity and/or Voltage
rating are used.
Power switch (-SPU2 version only)
Manual switch for power selection for the AvaSpec-SPU2
Left external power 12VDC, connect external power supply PS-12V/1.0A or 12 VDC batterypack
Middle - OFF
Right Power taken from USB bus, no additional power supply required
SMA entrance
Power LED
green
Scan LED yellow
/blue (BT)
12 VDC power supply
Synchronization connector
(only for multi-channel)
USB
connector
External IO
26 pole
Power Switch
(-SPU2 version only)
12 VDC OFF - USB power
Bluetooth
antenna conn.
(-BT only)