90 Apr-09
Avantes www.avantes.com info@avantes.com
Designations and Functions of Panel Controls
Front Panel
SMA-Connector / Protection Cap
The protection cap is only to avoid that the user could not unintentional look directly into the fiber
optic connector. The connector is for use only with SMA-Fiber connectors. Connect your Fiber optic
Cable first before starting the Deuterium Lamp. Please be sure to avoid direct radiation of skin
and eyes at the other fiber end.
Caution: This lamp produces ultra violet radiation, which can be harmful to the eyes. Lamps
must not be viewed directly without suitable approved eye protection being used.
OFF-TTL Shutter -Open
This switch is used to switch the lightsource on and operate TTL activated shutter (Normally closed)
Position switch
Lightsource
power
Shutter
OFF
OFF
Closed
TTL
ON
Closed or TTL activated the through backside DB-15
connector TTL signal, Pin 13 (High = Open, Low = Close)
ON
ON
Open
Deuterium both-Halogen
This switch is used to turn the deuterium and halogen light on.
Position switch
Deuterium
Halogen
Left
ON
OFF
Middle
ON
ON
Right
OFF
ON
Deuterium -both-Halogen