xi
Reagents
ISE SnapPak
TM
(BP5186) containing the following reagents:
Standard A
Use: For calibration of sodium, potassium, chloride, ionized calcium and
lithium in the AVL 9180 Electrolyte Analyzer
Contents: 350 mL
Active Ingredients: Na
+
150 mmol/L
K
+
5.0 mmol/L
Cl
-
115 mmol/L
Ca
++
0.9 mmol/L
Li
+
0.3 mmol/L
Additives: Germicides
Storage: Temperature: 5 - 30 °C (41 - 86 °F)
Stability: Expiration Date & Lot Number are printed on each container label.
Standard B
Use: For calibration of sodium, potassium, chloride, ionized calcium and
lithium in the AVL 9180 Electrolyte Analyzer
Contents: 85 mL
Active Ingredients: Na
+
100 mmol/L
K
+
1.8 mmol/L
Cl
-
72 mmol/L
Ca
++
1.5 mmol/L
Li
+
0.3 mmol/L
Additives: Germicides
Storage: Temperature: 5 - 30 °C (41 - 86 °F)
Stability: Expiration Date & Lot Number are printed on each container label.