Press [Output] to turn on the output.
Under this mode, it can monitor the current operation setting
address, left time and left cycle.
9. Remote Control
9.1.
With the remote control function provided, this series power
supply can communicate with PC by USB interface and en-
able all the panel operations by series port software.
9.2. USB Interface
The USB interface is installed on the real panel of instru-
ment.
10. Technical specifications
10.1. Data Sheet
The specifications apply when the power supply series are
powered on for at least 30 minutes under regulated tempera-
ture.
1. Model
2. Output range
3. Voltage
4. Current
5. OVP
6. OCP
Voltage output
Line regulation:≤0.01%+3mV
Load regulation:≤0.01%+3mV (I≤3A) / ≤0.02%+5mV
(I>3A)
Recovery time:≤100us (50% load change, minimum load
0.5A)
Ripple & Noise:≤1mV rms (I≤3A) (5Hz~1MHz) / ≤2mV
rms (I>3A) (5Hz~1MHz)
Temp. co-efficient:≤300ppm/°C
Setting accuracy:±(0.03% of reading + 10mV) (25±5°C)
Current output
Line regulation:≤0.2%+3mA
Load regulation:≤0.2%+3mA (I≤3A) / ≤0.2%+5mA (I>3A)
Ripple & Noise:≤3mA rms (I≤3A) / ≤6mA rms (I>3A)
Setting accuracy:±(0.2% of reading + 10mA) (25±5°C)
Display
Voltage: 5 digits display
Current: 5 digits display
Voltage resolution: 1mV
Current resolution: 0.1mA
Readback accuracy (25±5°C): ±(0.02% of reading +5mV) ;
±(0.02% of reading +5mA)
Protection: OLP, OVP, OCP, OTP and reverse polarity pro-
tection
Key lock:Provided
Interface:USB interface, SCPI commands provided
Save /Recall:100 sets
Insulati on :B et we en chassis and output terminal:
≥2 0MΩ/ 500VDC ; Betw een chassis and AC cord :
≥30MΩ/500VDC
Operation environment
Indoor use
Altitude: ≤2000m
Ambient temperature: 0~40°C
Relative humidity: ≤80%
Installation category: II
Pollution degree: 2
Storage environment: Ambient temperature: -10~70°C; Rel-
ative humidity: ≤70%
Power source: AC 110V/220V±10%, 50/60Hz
Accessories: Operation manual, Power cord, Software CD,
USB cable