Error messages and troubleshooting
Error message Cause What to do
There is a problem with automatic code detection. Repeat the test with a new test strip. If the problem persists,
contact the B. Braun representative in your area.
The ambient temperature is too high or too low. Put the meter in a location within the operating temperature range
(+10 °C to +40 °C / +50 °F to +104 °F) for at least 10 minutes and
repeat the test.
There is a problem with automatic code detection.
If the display alternates displaying Sun and Er6, avoid direct
sunlight entering the test strip port and repeat the test.
The Omnitest® 5 contains no readings. The memory for the measurement results is empty until you
have performed the first blood glucose test.