3015-9005 Rev 1 11
may operate the unit using the
supplied wall adapter.
!
NOTE: The sensor does not operate
when the low battery LED is on.
5. Check for sufficient airflow by pointing
the probe tip toward the floor,
covering it with your finger, then
releasing your finger. If proper flow
exists, the red ball should noticeably
rise up into the probe when you
uncover the probe tip. Note that the
actual height and final resting position
of the red ball are not important. If
the airflow ball does not rise:
a. Tap the probe lightly to ensure the
ball is not sticking.
b. Check the filter in the probe tip,
per Section 4.1. If the flow is still
insufficient, then the unit should
be sent for repair to the nearest
Authorized Service Center.