5.2 Replacing the Printed Circuit Board
Items required:
Medium flat-blade screwdriver
Medium Phillips screwdriver
Replacement printed circuit board (P/N 19-0418)
Procedure:
(See Figure 7)
1. Turn OFF instrument and lay it face down on work area.
2. Remove battery cover and batteries.
3. Remove sensor-matching resistor.
4. Remove four screws securing rear case.
5. Carefully lift rear case and flexible arm assembly away
from instrument, being careful not to pull wires attached to
printed circuit board. Lay rear case face down on work area.
6. Carefully remove printed circuit board from front case.
7. Unplug connectors J2 and J3 from printed circuit board.
8. Plug connectors J2 and J3 into new printed circuit board;
then install board into front case.
9. Install flexible arm assembly and rear case using the
screws that were removed in Step 4.
10. Replace sensor-matching resistor.
11. Replace batteries and battery cover.
12. Test instrument response (refer to Section 4.2).
Maintenance
12
Leakator 10
Instruction 19-9167

Other manuals for Bacharach Leakator 10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bacharach Leakator 10 and is the answer not in the manual?

Bacharach Leakator 10 Specifications

General IconGeneral
BrandBacharach
ModelLeakator 10
CategoryGas Detectors
LanguageEnglish

Related product manuals