EasyManuals Logo

berlinger Fridge-tag 2 E User Guide

Default Icon
2 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
In measurement mode, when no ALARM has been triggered, the
following information is indicated on the screen:
The OK symbol
The current date and time:16.09.2016 /13:47
The current temperature: +5.6 °C
Indication of an OK display
2
6101 2096
056
1 473
°C
Int.
Sensor
!
the ALARM symbol
an additional warning symbol
the corresponding arrow: upper/lower alarm
the current date and time:16.09.2016 /15:42
the current temperature: +8.8 °C
In measurement mode, when an ALARM has been triggered, the
following information is indicated on the screen:
Indication of an ALARM display
2
6101 2096
088
1 425
°C
Int.
Sensor
!
!
!
!
!
2
55
0 000
max.
°C
Int.
Sensor
the OK symbol
the corresponding arrow
(example: high limit records of the current day today)
the highest recorded temperature of that day:
+5 .5 ° C
the time duration out of the preset temperature
high limit: 00:00 (hrs: min)
The following information is indicated on the display:
Reading the history
Option 1:
Read out day-per-day directly on the
device (30 day history)
Example of an OK display
!
Press
READ
once
!
2
6101 2099
1 218
!
1st displayed screen
Press
READ
The following information is indicated on the display:
the ALARM symbol
the corresponding arrow
(example: Low limit records and the day of ALARM)
the date of alarm:1 9.09.2016
the time of alarm:18:21
Example of an ALARM display
!
!
Reading the display during measurement
2
-
011
0 351
min.
Int.
Sensor
!
Press READ
a second time
The following additional information is indicated on the display:
Note: Continue repetitively pressing the READ button to read
out the details of the past 30 days.
Lowest recorded temperature of that day: -1. 1 ° C
The time duration out of the preset temperature
low limit: 01:35 (hrs:min)
2nd displayed screen
!
In addition to the ALARM symbol
each newly triggered alarm is
indicated by the warning symbol.
!
Option 2: The warning and ALARM symbol (X)
will remain visible until the user reads the details of
the triggered alarm/s from the display. After that both
symbols will disappear and the display will go back to
the OK Symbol ( ).
The warning symbol
Option 1: The warning and ALARM symbol
(X) will remain visible until the user reads the
details of the triggered alarm/s from the display.
After that the warning symbol will disappear.
The ALARM symbol (X) cannot be changed.

Other manuals for berlinger Fridge-tag 2 E

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the berlinger Fridge-tag 2 E and is the answer not in the manual?

berlinger Fridge-tag 2 E Specifications

General IconGeneral
Brandberlinger
ModelFridge-tag 2 E
CategoryMeasuring Instruments
LanguageEnglish