PRODUCT ILLUSTRATION
124mm
320mm
INSTANT FLEXIBLE DIGITAL THERMOMETER
OWNER S M ANUAL
Model:DMT-4318
Warning :
Read instructions thoroughly before using digital thermometer.
Choking Hazard: Thermometer cap and battery may be fatal if swallowed. Do not allow children
to use this device without parental supervision.The Patient is an intended operator.
Do not use thermometer in ear. Designed use is for oral, rectal, and armpit (axilla) readings only.
Do not place thermometer battery near extreme heat as it may explode.
Remove battery from the device when not in operation for a long time.
The use of temperature readings for self-diagnosis is dangerous. Consult your doctor for the
interpretation of results. Self-diagnosis may lead to the worsening of existing disease conditions.
Do not attempt measurements when the thermometer is wet as inaccurate readings may result.
Do not bite the thermometer. Doing so may lead to breakage and/or injury.
Do not attempt to disassemble or repair the thermometer. Doing so may result in inaccurate readings.
After each use, disinfect the thermometer especially in case the device is used by more than one person.
Do not force the thermometer into the rectum. Stop insertion and abort the measurement when pain is
present. Failure to do so may lead to injury.
Do not use thermometer orally after being used rectally.
For children who are two years old or younger, please do not use the devices orally.
If the unit has been stored at temperatures over 41104(540℃), leave it in 41104
(540℃) ambient temperature for about 15 minutes before using it.
Figure 2
2. Position thermometer in desired location (mouth, rectum, or armpit.)
a) Oral Use: Place thermometer under tongue as indicated by
position shown in Figure 2. Close your mouth and breathe evenly
through the nose to prevent the measurement from being influenced
by inhaled/exhaled air.
b) Rectal Use: Lubricate silver probe tip with petroleum jelly for easy insertion. Gently insert
sensor approximately 1cm (less than 1/2 ") into rectum.
c) Armpit Use: Wipe armpit dry. Place probe in armpit and keep arm pressed firmly at side.
From a medical viewpoint, this method will always provide inaccurate readings, and should not
be used if precise measurements are required.
3. The degree sign flashes throughout the testing process. When flashing stops an alarm will beep for
approximately 10 seconds. The measured reading will appear on the LCD simultaneously. The
minimum measurement time until the signaling tone (beep) must be maintained without exception. The
measurement continues even after the buzzer notification. So that in order to achieve better body
temperature measurement result, recommend to keep the probe in mouth and rectum about 2 minutes,
or in armpit about 5 minutes regardless of the beep sound and at least 30 seconds measurement interval
should be maintained.
4. To prolong battery life, press the On/Off Button to turn unit off after testing is complete. If no action
is taken, the unit will automatically shut off after around10 minutes.
*Note: Normally Alarm b eeps m ore ra pidly w hen te mperature the buzzes are Bi-Bi-Bi- B i-;
reaches 100 (37.8 ) or higher ,and t he buzzes are Bi-Bi-Bi----- Bi-Bi-Bi------- Bi-Bi-Bi
CLEANING AND DISINFECTION
1)Immerse the thermometer probe in distilled water for at least 1 minute;
2)Using a clean, soft cloth to wipe the thermometer down to remove any residue;
3)Repeat step 1 and 2 for three times until no soil is seen with visual inspection after cleaning;
4)For thoroughly clean and disinfection, please use method A or B
Method A(High level disinfection): immerse the thermometer probe in 0.55% OPA(O-Phthaldehyde),
such as CIDEX OPA, for at least 12 minutes under temperature at 20℃;
Method B(Low level disinfection): Using a clean soft cloth dipped in 70% medical alcohol,wipe the probe
3 times,at least one minute for each time.
5)Repeat step 1 to 3 to remove OPA residuals;
Note1: Rectal use is not recommended for home use as OPA will not be readily available outside of a hospital.
If rectal measurement is necessary, we strongly recommend high level disinfection.
Note2: Please operate according to the manual of OPA for reference.
To prevent damage to the thermometer please note and observe the following:
-Do not use benzene, paint thinner, gasoline or other strong solvents to clean the thermometer.
-Do not attempt to disinfect the sensing probe (tip) of the thermometer by immersing in alcohol, OPA
or in hot water (water over 122(50) for long time.
-Do not use ultrasonic washing to clean the thermometer.
Figure 4 Figure 5Figure 3
BATTERY REPLACEMENT
1. Replace battery when appears in the lower right corner of LCD display.
2. Pull battery cover off as shown in Figure 3.
3. Gently pull out plastic circuit board with battery chamber approximately 1 cm (slightly less than1/2 ".)
(See Figure 4)
4. Use pointed object such as a pen to remove old battery. Discard battery lawfully. Replace with new
1.5V DC button type LR41 or SR41,UCC392, or equivalent. Be sure battery is installed with +
polarity facing up. (See Figure 5)
5. Slide battery chamber back into place and attach cover.
TROUBLESHOOTING
Problem
Solution
Error message
The system is not
functioning properly.
Unload the battery, wait for 1 minute and repower
it. If the message reappears, contact the retailer
for service.
Temperature taken is
lower than 90.0°F(32.0°C)
Turn off, wait one minute and take a new temperature
via close contact and sufficient rest.
Temperature taken is
higher than 109.9°F(42.9°C)
Dead battery: Battery icon is
flashing, cant be measurable.
Replace the battery.
Turn off, wait one minute and take a new temperature
via close contact and sufficient rest.
LIMITED WARRANTY
The thermometer is guaranteed for one year from the date of purchase. If the thermometer does not
function properly due to defective components or poor workmanship, we will repair or replace it free of
charge. All components are covered by this warranty excluding the battery. The warranty does not cover
damages to your thermometer due to improper handling. To obtain warranty service, an original or copy
of the sales receipt from the original retailer is required.
Disposal of this product and used batteries should be carried out
in accordance with the national regulations for the disposal of
electronic products.
FCC INFORMATION
Caution: Changes or modifications to this unit not expressly approved by the party responsible for
compliance could void the user authority to operate the equipment.
*Note:
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment generates, uses, and can radiate radio
frequency energy. If this equipment does cause harmful interference to radio or television reception,
which can be determined by turning the equipment off and on, the user is encouraged to try and correct
the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the distance between the equipment and the receiver.
- Connect the equipment to an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
Made in China
Document No. :JDMT-2104-002
Article No.: 001
Version: A
Date of Issue: 2020.01
CALIBRATION
The thermometer is initially calibrated at the time of manufacture. If the thermometer is used according to
the use instruction, periodic readjustment is not required. However, we recommend checking calibration
every two years or whenever clinical accuracy of the thermometer is in question. Turn on the thermometer
and insert into the water bath and then check the laboratory accuracy. Please send the complete device to the
dealers or manufacturer. ASTM laboratory accuracy requirements in the display range of 98.6 to 102.2 °F
(37.0 to 39.0 °C) for electronic thermometers is ±0.2°F(±0.1°C).
The above recommendations do not supersede the legal requirements. The user must always comply with
legal requirements for the control of the measurement, functionality, and accuracy of the device which
are required by the scope of relevant laws, directives or ordinances where the device is used.
JOYTECH HEALTHCARE CO.LTD.
No.365, Wuzhou Road, Yuhang Economic Development Zone, Hangzhou city, 311100 Zhejiang,China
Battery Cap
On/Off Button
LCD
Probe
Figure 1
DIRECTIONS
1. Press the On/Off Button next to LCD display. A tone will sound as the screen shows ,
followed by last recorded temperature. After showing the self-test temperature, the thermometer is
now in the testing mode.
Measure Range:
Accuracy:
Operating mode: Direct Mode
Display:
Memory:
Battery:
Battery life:
Dimension:
Weight:
90.0-109.9(32.0℃-42.9 ) ( / c h osen b y m anufacturer )
±0.2(±0.1) during 95.9~107.6(35.5~42.0) at
64.4~82.4(18~28 ) ambient operating range
±0.4(±0.2) for other measuring and ambient operating range
Liquid crystal display, 3 1/2 digits
For storing the last measured value
One 1.5 V DC. button battery (size LR41or SR41, UCC 392)
13.9cm x 2.2cm x 1.3cm (L x W x H)
Approx. 11 grams including battery
Type:
Digital Thermometer (Not Predictive)
PRECAUTION
*The performance of the device may be degraded should one or more of the following occur:
- Operation outside the manufacturers s tated te mperature and humidity range.
- Storage outside the manufacturers s tated te mperature and humidity range.
- Mechanical shock (for example, drop test) or degraded sensor.
- Patient temperature is below ambient temperature.
*Portable and mobile RF communications can affect the device. The device needs special pre-cautions
regarding EMC according to the EMC information provided in the accompany documents.
*Do not use the devices in the MR environment.
CONTENTS
1 Thermometer, 1 Owner s Manual, 1 Storage Case
SPECIFICATIONS
SYMBOL EXPLANATION
Direct Current
Type BF Applied Part
Batch Code
Manufacturer
PLEASE READ CAREFULLY BEFORE USING
This digital thermometer provides a quick and highly accurate reading of an individual's body
temperature. To better understand its functions and to provide years of dependable
results, please read all instructions first.
This appliance conforms to the following standards:
ASTM E1112 Standard Specification for Electronic Thermometer for Intermittent Determination of
Patient Temperature,
ISO 80601-2-56 Medical electrical equipment —Part 2-56:Particular requirements for basic safety
and essential performance of clinical thermometers for body temperature measurement,
IEC 60601-1-11 Medical electrical equipment —Part 1-11: General requirements for basic safety and
essential performance –Collateral Standard: Requirements for medical electrical equipment and
medical electrical systems used in the home healthcare environment and complies with the
requirements of IEC 60601-1-2(EMC), AAMI/ANSI ES60601-1(Safety) standards. And the
manufacturer is ISO 13485 certified.
Consult Accompanying Documents
Storage and Transportation
Temperature Limit:-4 131
(-20~55℃)
-2 0
55
Approx. 200hours of continuous operation or 1 year with 3
measurements per day
Ambient operating range:
Storage and transportation
condition:
Temperature: 41104(540)
Relative humidity: 15%~95%RH
Atmospheric Pressure : 700hPa ~ 1060hPa
Temperature:-4℉~131 (-20~55)
Relative humidity: 15%~95%RH
Atmospheric Pressure : 700hPa ~ 1060hPa
Classification:
Type BF
Ingress Protection Rating:
IP 27
Expected service life: Three years
/ SWITCHABLE
Temperature readings are available in the Fahrenheit or Celsius scale (/; located in the upper right
corner of LCD.) With the unit off, press and hold the On/Off Button for approximately 2 seconds to
change the current setting.
The digital thermometers are intended to measure the human body temperature in regular
mode orally, rectally or under the arm. And the devices are reusable for clinical or home use on people of
all ages, including children under 8 years old with adult supervision.
Indications For Use
Question and Answer IconNeed help?

Do you have a question about the BestMed DMT-4318 and is the answer not in the manual?

Summary

Safety Warnings and Precautions

Choking Hazard and Parental Supervision

Warning about choking hazard from cap and battery; parental supervision required.

Intended Use and Temperature Measurement Sites

Device is for oral, rectal, and armpit readings only. Do not use in ear.

Battery Safety Precautions

Do not place battery near extreme heat or explode. Remove if not in use.

Danger of Self-Diagnosis

Self-diagnosis is dangerous; consult doctor for results and potential worsening.

Measurement Integrity and Maintenance

Avoid wet measurements, biting, disassembly. Disinfect after each use.

Rectal Use Caution and Pain Management

Do not force rectal insertion. Stop if pain is present. Do not use orally after rectal use.

Temperature Acclimation Before Use

If stored below 41°F, leave in 41°F-104°F ambient for 15 mins before use.

Indications and Standards

Intended Use for Temperature Measurement

Measures body temperature orally, rectally, or under the arm for clinical/home use.

Standards Conformance and Certification

Appliance conforms to ASTM E1112, ISO 80601-2-56, IEC 60601-1-11 standards.

Package Contents and Illustration

Included Items in Package

Package includes 1 Thermometer, 1 Owner's Manual, 1 Storage Case.

Thermometer Component Diagram

Illustration shows Probe, LCD, Battery Cap, and On/Off Button.

Operational Precautions

Factors Affecting Device Performance

Performance may degrade due to environmental range, shock, or sensor issues.

RF Communication Interference and EMC

Portable RF devices can affect performance; requires EMC precautions.

MR Environment Restriction

Do not use the device in an MR environment.

Technical Specifications

Device Type and Measurement Range

Digital thermometer with a measure range of 90.0°F-109.9°F (32.0°C-42.9°C).

Measurement Accuracy

Accuracy is ±0.2°F/±0.1°C within specific ambient ranges, ±0.4°F/±0.2°C otherwise.

Operating Mode and Display Features

Operates in Direct Mode with a liquid crystal display showing 3 1/2 digits.

Memory and Battery Information

Stores last measured value. Uses one 1.5V DC button battery (LR41/SR41).

Battery Life and Physical Dimensions

Approx. 200 hours battery life. Dimensions: 13.9cm x 2.2cm x 1.3cm.

Ambient and Storage Conditions

Specifies operating ambient temperature, humidity, and pressure ranges.

Ingress Protection and Classification

Ingress Protection Rating IP 27, Classification Type BF.

Operating Instructions

Temperature Unit Setting

Press and hold On/Off Button for 2 seconds to change °F/°C setting.

Thermometer Activation

Press On/Off Button; unit shows 188.8. and last recorded temperature.

Measurement Procedures

Positioning for Measurement

Instructions for oral, rectal, and armpit positioning for accurate readings.

Measurement Duration and Beeping

Maintain measurement until beep stops. Alarm beeps faster at higher temperatures.

Recommended Measurement Time

Recommend keeping probe in mouth/rectum for 2 mins, armpit for 5 mins.

Automatic Shut-off

Unit automatically shuts off after approximately 10 minutes of inactivity.

Troubleshooting Common Errors

Error Message Lo or Hi

Indicates temperature is too low or too high; advises re-measurement.

Error Message Err

System not functioning properly; advises battery reset or contacting retailer.

Dead Battery Indicator

Battery icon flashing indicates dead battery; requires replacement.

Battery Replacement Guide

When to Replace Battery

Replace battery when the battery icon appears in the lower right corner.

Battery Replacement Steps

Detailed steps for opening cover, removing old battery, and inserting new one.

Cleaning and Disinfection Procedures

Basic Cleaning Steps

Immerse probe in distilled water, wipe with soft cloth, repeat for thorough cleaning.

High-Level Disinfection (OPA)

Immerse probe in 0.55% OPA for 12 mins at 20°C for high-level disinfection.

Low-Level Disinfection (Alcohol)

Wipe probe 3 times with 70% medical alcohol for low-level disinfection.

Cleaning Precautions

Avoid harsh solvents, hot water over 50°C, or ultrasonic washing.

Calibration and Maintenance

Initial Calibration

Thermometer is initially calibrated at manufacture; no periodic readjustment needed.

Calibration Check Recommendation

Recommend checking calibration every two years or if accuracy is questioned.

Regulatory Information

FCC Information and Interference

Class B digital device compliance with FCC Rules; guidance on correcting interference.

Warranty and Disposal

Limited Warranty Details

One-year warranty covers defects; excludes battery and improper handling.

Product and Battery Disposal

Dispose of product and batteries according to national regulations.

Overview

This document describes the BESTMED® INSTANT FLEXIBLE DIGITAL THERMOMETER, Model: DMT-4318, an owner's manual detailing its function, specifications, usage, and maintenance.

Function Description:

The BESTMED® INSTANT FLEXIBLE DIGITAL THERMOMETER is designed to measure human body temperature in regular mode orally, rectally, or under the arm (axilla). It provides quick and highly accurate readings and is reusable for clinical or home use on people of all ages, including children under 8 years old with adult supervision. The device conforms to several standards, including ASTM E1112 for electronic thermometers, ISO 80601-2-56 for clinical thermometers, IEC 60601-1-11 for home healthcare environments, IEC 60601-1-2 (EMC), AAMI/ANSI ES60601-1 (Safety), and is ISO 13485 certified.

Important Technical Specifications:

  • Type: Digital Thermometer (Not Predictive).
  • Measure Range: 90.0°F - 109.9°F (32.0°C - 42.9°C). The user can choose between °C/°F.
  • Accuracy:
    • ±0.2°F (±0.1°C) during 95.9°F - 107.6°F (35.5°C - 42.0°C) at 64.4°F - 82.4°F (18°C - 28°C) ambient operating range.
    • ±0.4°F (±0.2°C) for other measuring and ambient operating ranges.
  • Operating Mode: Direct Mode.
  • Display: Liquid crystal display, 3 1/2 digits.
  • Memory: Stores the last measured value.
  • Battery: One 1.5 V DC button battery (size LR41 or SR41, UCC 392).
  • Battery Life: Approximately 200 hours of continuous operation or 1 year with 3 measurements per day.
  • Dimension: 13.9cm x 2.2cm x 1.3cm (L x W x H).
  • Weight: Approximately 11 grams including battery.
  • Expected Service Life: Three years.
  • Ambient Operating Range:
    • Temperature: 41°F - 104°F (5°C - 40°C).
    • Relative Humidity: 15% - 95% RH.
    • Atmospheric Pressure: 700hPa - 1060hPa.
  • Storage and Transportation Condition:
    • Temperature: -4°F - 131°F (-20°C - 55°C).
    • Relative Humidity: 15% - 95% RH.
    • Atmospheric Pressure: 700hPa - 1060hPa.
  • Ingress Protection Rating: IP 27.
  • Classification: Type BF.

Usage Features:

  • °F/°C Switchable: Temperature readings are available in Fahrenheit or Celsius. To change the setting, with the unit off, press and hold the On/Off Button for approximately 2 seconds.
  • Operation:
    1. Press the On/Off Button. A tone will sound, the screen will show "188.8.", followed by the last recorded temperature. After showing the self-test temperature, the thermometer enters testing mode.
    2. Positioning:
      • Oral Use: Place the thermometer under the tongue as indicated in Figure 2. Close the mouth and breathe evenly through the nose to prevent measurement influence from inhaled/exhaled air.
      • Rectal Use: Lubricate the silver probe tip with petroleum jelly and gently insert the sensor approximately 1cm (less than 1/2") into the rectum.
      • Armpit Use: Wipe the armpit dry, place the probe in the armpit, and keep the arm pressed firmly at the side. This method may provide inaccurate readings and should not be used if precise measurements are required.
    3. Measurement Process: The degree sign flashes during testing. When flashing stops, an alarm will beep for approximately 10 seconds, and the measured reading will appear on the LCD. The minimum measurement time until the signaling tone (beep) must be maintained. Measurement continues after the buzzer notification. For better accuracy, it is recommended to keep the probe in the mouth and rectum for about 2 minutes, or in the armpit for about 5 minutes, regardless of the beep sound, and maintain at least a 30-second measurement interval.
    4. Alarm Tones: Normal buzzes are "Bi-Bi-Bi-Bi-". Alarm beeps more rapidly ("Bi-Bi-Bi----- Bi-Bi-Bi------- Bi-Bi-Bi") when the temperature reaches 100°F (37.8°C) or higher.
    5. Auto Shut-off: To prolong battery life, press the On/Off Button to turn the unit off after testing. If no action is taken, the unit will automatically shut off after approximately 10 minutes.
  • Troubleshooting:
    • "Lo" message: Temperature taken is lower than 90.0°F (32.0°C). Solution: Turn off, wait one minute, and take a new temperature via close contact and sufficient rest.
    • "Hi" message: Temperature taken is higher than 109.9°F (42.9°C). Solution: Turn off, wait one minute, and take a new temperature via close contact and sufficient rest.
    • "Err" message: The system is not functioning properly. Solution: Unload the battery, wait 1 minute, and repower it. If the message reappears, contact the retailer for service.
    • Flashing battery icon: Dead battery, cannot be measurable. Solution: Replace the battery.

Maintenance Features:

  • Battery Replacement:
    1. Replace the battery when the battery icon appears in the lower right corner of the LCD display.
    2. Pull off the battery cover (Figure 3).
    3. Gently pull out the plastic circuit board with the battery chamber approximately 1 cm (Figure 4).
    4. Use a pointed object (e.g., a pen) to remove the old battery. Discard lawfully.
    5. Replace with a new 1.5V DC button type LR41 or SR41, UCC392, or equivalent. Ensure the battery is installed with "+" polarity facing up (Figure 5).
    6. Slide the battery chamber back into place and attach the cover.
  • Cleaning and Disinfection:
    1. Immerse the thermometer probe in distilled water for at least 1 minute.
    2. Use a clean, soft cloth to wipe the thermometer down to remove any residue.
    3. Repeat steps 1 and 2 three times until no soil is visible.
    4. For thorough cleaning and disinfection, use Method A or B:
      • Method A (High-level disinfection): Immerse the thermometer probe in 0.55% OPA (O-Phthaldehyde), such as CIDEX OPA, for at least 12 minutes under temperature at 20°C. (Note: Rectal use is not recommended for home use as OPA is not readily available outside of a hospital. If rectal measurement is necessary, high-level disinfection is strongly recommended.)
      • Method B (Low-level disinfection): Use a clean soft cloth dipped in 70% medical alcohol to wipe the probe 3 times, at least one minute for each time.
    5. Repeat steps 1 to 3 to remove OPA residuals.
    6. Important Notes for Cleaning:
      • Operate according to the OPA manual for reference.
      • Do not use benzene, paint thinner, gasoline, or other strong solvents to clean the thermometer.
      • Do not attempt to disinfect the sensing probe (tip) by immersing it in alcohol, OPA, or hot water (over 122°F (50°C)) for a long time.
      • Do not use ultrasonic washing to clean the thermometer.
  • Calibration:
    • The thermometer is initially calibrated at the time of manufacture. Periodic readjustment is not required if used according to instructions.
    • However, it is recommended to check calibration every two years or whenever clinical accuracy is in question.
    • To check, turn on the thermometer, insert it into a water bath, and then check the laboratory accuracy.
    • Send the complete device to the dealers or manufacturer if needed.
    • ASTM laboratory accuracy requirements in the display range of 98.6 to 102.2 °F (37.0 to 39.0 °C) for electronic thermometers is ±0.2°F (±0.1°C).
    • These recommendations do not supersede legal requirements; users must comply with legal requirements for measurement control, functionality, and accuracy.
  • Limited Warranty: The thermometer is guaranteed for one year from the date of purchase. It covers defective components or poor workmanship, excluding the battery. Damages due to improper handling are not covered. To obtain warranty service, an original or copy of the sales receipt from the original retailer is required.
  • Disposal: This product and used batteries should be disposed of in accordance with national regulations for electronic products.