MODE Selection Key Function
Pressing this button also enables access to the Emissivity set up function
(EMS),activation/de-activation of the continuous measurement mode, of the
activation/de-activation and regulation of the high (HAL) and low (LAL)
alarm. Each time you press the selection MODE key, the instrument passes
to the subsequent function. The diagram shows the cycle sequence of the
various
modes.
EMS adjustment.
The Emissivity(EMS) is digitally adjustable from 0.10 to 1.0.
Activation/De-activation mode for CONTINUOUS
MEASUREMENT
The continuous measurement mode is particularly useful for the continuous
monitoring of temperature. Press the “up” and “down”keys to activate it/de-
activate it. Press the measurement push button (trigger) to conrm continuous
measurement mode.
The instrument continues to display temperature even if
the trigger is not pressed again.
In the continuous measurement mode, press the up and down keys to adjust
emissivity or activate/de-activate the high (HAL) low (LAL) alarm. Press the
measurement button (trigger) to conrm the High/Low alarm method.
High/low Alarm Regulation
from
-50 to 800
o
C(-58℉ ~ 1472℉)
EMS
Lock
on/off
Hal
on/off
HAL
Low
on/off
Low
MODE