EasyManua.ls Logo

Beurer BM 50 - Troubleshooting Guide; Specifications

Beurer BM 50
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
11. Troubleshooting Guide
In case of errors, “ _” will appear in the LCD Display, where “_” is a number 1-6. Error messages may
appear if:
No pulse rate detected ( )
You speak or move during measurement ( )
The cuff was not attached correctly or the cuff is fastened too tightly or loosely and measurement
takes longer than 25 seconds (
)
A system or device error occurs ( )
Air pressure is higher than 300 mmHg ( )
Batteries are almost depleted ( )
In such cases, repeat the measurement. Do not move or speak. If necessary, replace the batteries.
12. Specifications
Type BM 50
Measurement Method Oscillometric, non-invasive blood pressure measurement on the upper arm
Measurement Range
Cuff Pressure 0 – 300 mmHg
Systolic 50 – 250 mmHg
Diastolic 30 – 200 mmHg
Pulse 30 180 beats per minute
Display Accuracy
Pressure
Pulse
± 3 mmHg
± 5%
Dimensions 5.3 in L x 4.0 in W x 2.1 in H (134 mm L x 102 mm W x 53 mm H)
Weight 14.32 oz (406 g) with cuff, without batteries
Cuff Size 8.716.5 in (2242 cm)
Operating Range 41104 °F (5+40 °C), 1590% relative humidity (non-condensing)
Storage Range -13158 °F (-25+70 °C), ≤ 90% relative humidity
Power Supply 4 x 1.5V
AA batteries
Battery Life
Approx. 250 measurements, depending on the blood pressure level and/
or pump pressure
The serial number is located on the device or in the battery compartment. These specifications are subject
to change without notice for purpose of improvement.
This unit is in line with European Standard EN 60601-1-2 (In accordance with CISPR 11,
IEC 61000-3-2, IEC 61000-3-3, IEC 61000-4-2, IEC 61000-4-3, IEC 61000-4-4, IEC 61000-4-5,
IEC 61000-4-6, IEC 61000-4-8 and IEC 61000-4-11) and is subject to particular precautions with regard
to electromagnetic compatibility (EMC). Please note that portable and mobile HF communication
systems may interfere with this unit.
This device is in line with the EU Medical Devices Directive 93/42/EEC, the “Medizinproduktegesetz”
(German Medical Devices Act) and the standards EN ISO 81060-1 (non-invasive sphygmomanometers,
Part 1: Requirements and test methods), EN 1060-3 (non-invasive sphygmomanometers, Part 3:
Supplementary requirements for electro-mechanical blood pressure measuring systems) and IEC
80601-2-30 (Medical electrical equipment – Part 2–30: Particular requirements for the safety and
essential performance of automated noninvasive blood pressure monitors).

Related product manuals