27
Operating
Temperature
and humidity
41 °F to 104 °F (+ 5 °C to + 40 °C),
≤ 90 % relative humidity
Transport
and storage
Temperature
and humidity
- 4 °F to 131 °F(- 20 °C to + 55 °C),
≤ 95 % relative humidity
Power Supply
4 x 1.5 V
AA batteries (alkaline type
LR6)
Battery Life for approx. 200 measurements, depending on
blood pressure level or inflation pressure
Accessories
with storage pouch, instruction manual,
4 pcs “AA” battery LR6, standard and large
cuff, USB connection cable, AC adapter
Classification
internal power supply, Type BF applied part,
IPX0, no AP or APG, continuous operation
Auto shut-off approx. 1 minute after last button operation
These specifications are subject to change without notice
for purpose of improvement.
14. Mains part
Model No.
FW 7575M/US/6/06
Input
100 – 240 V, 50 – 60 Hz
Output
6 V DC, 600 mA, only in connection with
beurer blood pressure monitor
Supplier
Friwo Gerätebau GmbH
Protection
This device is double insulated and
protected against short circuit and
overload by a primary thermal fuse.
Make sure to take the batteries out of the
compartment before using the mains part.
Double insulated / equipment class 2
Polarity of the the DC voltage connec-
tion
Enclosures and
Protective Covers
Equipment enclosed to protect against con-
tact with live parts, and with parts which
can become live (finger, pin, hook test).
The operator shall not contact the patient
and the output plug of mains part simulta-
neously.
ELECTROMAGNETIC COMPATIBILITY
This unit is in line with Standard IEC 60601-1-2 and is
subject to particular precautions with regard to electro-
magnetic compatibility (EMC). Please note that portable
and mobile HF communication systems may interfere with
this unit. More details can be requested from the stated
Customer Service address or found at the end of the
instructions for use.