15
Read these instructions for use carefully and keep them for later
use, be sure to make them accessible to other users and observe
the information they contain.
ENGLISH
Table of contents
1. Getting to know your device ....................................15
2. Included in delivery ...................................................16
3. Signs and symbols ....................................................16
4. Intended use ..............................................................17
5. Notes ..........................................................................17
6. Device description ....................................................19
7. Initial use ....................................................................20
7.1 Assembling the breast pump .................................20
7.2 Cleaning and sterilising the breast pump ............... 21
8. Operation .................................................................... 22
8.1 Tips to ensure a better milk flow ............................22
8.2 Operating the breast pump ....................................22
8.3 Storing the breast milk ...........................................23
8.4 Feeding with the breast milk ..................................23
9. What if there are problems? .....................................24
9.1 Replacement parts and wearing parts ...................24
10. Disposal ................................................................... 25
11. Technical data .........................................................25
12. Mains part ................................................................26
13. Warranty/service .....................................................26
Dear customer,
Thank you for choosing one of our products. Our name stands for high-quality, thoroughly tested prod-
ucts for applications in the areas of heat, gentle therapy, blood pressure/diagnosis, weight, massage,
beauty, air and babycare. Read these instructions for use carefully and keep them for later use, be sure
to make them accessible to other users and observe the information they contain.
With kind regards,
Your Beurer team
1. Getting to know your device
The Beurer BY 40 breast pump allows nursing mothers to express their breast milk easily and conveniently.
If not enough breast milk is taken from mothers who have recently given birth, this can lead to the pain-
ful build-up of milk, which in the worst-case scenario could cause inflammation of the nipples (mastitis).
This build-up of milk can be prevented through targeted expressing of breast milk using the breast pump.
In addition, breast pumps help to alleviate cracked or sore nipples. You can use the integrated stimulation
function to draw out flat or inverted nipples, making it easier to then pump the breast milk. The breast
pump also allows you to store breast milk in the bottle, for example if you will be travelling or away from
your baby for an extended period of time.
Babies who cannot be breast-fed directly for health reasons (due to a premature birth or cleft palate) can
still be given breast milk by using the breast pump.

Other manuals for Beurer BY 40

Question and Answer IconNeed help?

Do you have a question about the Beurer BY 40 and is the answer not in the manual?

Overview

The Beurer BY 40 breast pump is designed for nursing mothers to express breast milk easily and conveniently. It helps prevent the painful build-up of milk that can lead to mastitis, and can alleviate cracked or sore nipples. The integrated stimulation function helps draw out flat or inverted nipples, making pumping easier. The device also facilitates storing breast milk in a bottle for travel or extended periods away from the baby. It allows babies who cannot be breast-fed directly due to health reasons (e.g., premature birth, cleft palate) to still receive breast milk.

Included in delivery:

  • Breast pump (motor unit)
  • Silicone tube
  • Pump attachment with silicone membrane and silicone valve
  • Silicone cushion (standard size)
  • Bottle (180ml)
  • Screw plug
  • 2 x adapter for Avent and NUK bottles
  • Mains plug
  • 4 x 1.5 V AA batteries
  • Instructions for use

Important technical specifications:

  • Model/Type: BY 40
  • Power supply: 4 x 1.5V AA batteries (Alkaline batteries) or mains plug
  • Voltage (input): 100-240V (AC~) 50/60 Hz
  • Voltage (output): 6V (DC) 1.0 A
  • Dimensions: 168 x 100 x 53 mm
  • Weight: 310 g (without batteries)
  • Maximum suction pressure: 0.32 bar
  • Permissible operating conditions: +5°C to +40°C, 30% to 85% relative humidity (non-condensing), 700 – 1060 hPa ambient pressure
  • Permissible storage conditions: -20°C to +60°C, 10% to 95% relative humidity (non-condensing), 700 – 1060 hPa ambient pressure
  • Expected life: 500 hours
  • Equipment for intermittent use: 30 min. ON/30 min. OFF.
  • Safety requirements: In accordance with EN 60601-1
  • Child use and care articles: Drinking equipment in accordance to EN 14350
  • Unit for suction: Breast milk
  • Unit for medium vacuum and low flow
  • Class II equipment (safety insulation)
  • Unit suitable for use in the absence of atmosphere containing oxidising, exploding gases, nitrous oxide.
  • BF type equipment (unit with specific grade of protection against electrical hazards)
  • Mains part (Model no. LXCP12-06): Input 100-240 V, 50-60 Hz 0.3 A max; Output 6 V DC, 1 A. Insulated/protection class 2.

Usage features:

The device features a display and buttons for operation.

  • Battery status: Indicated for operation without mains plug.
  • Pumping mode: To pump breast milk.
  • Time display: Timer for stimulation/duration of use for pumping mode.
  • Stimulation or pumping level: Adjustable.
  • Stimulation mode: To stimulate the nipple – no breast milk is pumped.
  • + button: To increase stimulation or pumping level.
  • ON/OFF button: To switch the device on or off.
  • – button: To decrease stimulation or pumping level.
  • MODE button: To change between pumping and stimulation mode.

Initial Use:

  1. Assembly: Screw the pump attachment onto the bottle, ensuring it is firmly secured. Insert the silicone tube into the pump attachment and the breast pump.
  2. Power: Connect the breast pump to the mains cable and insert the mains plug into the socket. Alternatively, for battery operation, open the battery compartment on the rear side of the motor unit, insert four AA batteries (included), ensuring correct polarity, and close the compartment. A single battery charge with new AA batteries lasts approximately 60 minutes at the lowest setting.
  3. Operation:
    • Wash hands and breasts thoroughly before use.
    • Sit with your upper body slightly forward, using a pillow for back support.
    • Place the breast shield with silicone cushion gently on the breast, ensuring the nipple is completely enclosed and no air pockets. The breast shield must be drawn firmly onto the breast for proper function.
    • Press the ON/OFF button to switch on. The breast pump automatically starts in stimulation mode (indicated by ◎) with a two-minute timer. Adjust stimulation intensity using the +/– buttons.
    • After two minutes, the pumping mode indicator will flash. Press the MODE button to begin pumping breast milk. Select a comfortable suction intensity using the +/– buttons.
    • If you wish to switch back to stimulation mode, press the MODE button; the stimulation and two-minute timer will restart.
    • The breast pump has a memory function, recalling the last programmed stimulation or pumping level when switched on.
    • To switch off, press the ON/OFF button.

Tips for better milk flow:

  • Choose a relaxing location and time for pumping.
  • Having your child nearby or looking at their picture can enhance milk flow.
  • If feeding from one breast, use the pump on the other simultaneously.
  • Warmth and relaxation encourage milk flow; pump after a shower or bath.
  • For swollen or sore breasts, apply a warm towel for a few minutes to improve milk flow.
  • If milk doesn't flow immediately, relax and try again later.
  • If no milk is extracted after five minutes, stop and try again later.
  • Consult a doctor if you experience sharp pains.

Storing breast milk:

  • Ensure the breast pump and container are disinfected before use.
  • Store breast milk in disinfected containers immediately after expressing; do not leave it at room temperature.
  • If freezing, mark the bottle with the date of expressing.
  • Storage periods:
    • Room temperature: Can be kept for 6 hours.
    • Refrigerator (approx. 3-5°C): Can be kept for 24 hours (do not store in the fridge door).
    • Freezer (approx. -16°C): Can be kept for 3 months.
    • Defrosted breast milk: Use immediately; can be kept for 10 hours in the refrigerator. Do not refreeze.

Feeding with breast milk:

  • Disinfect the bottle before filling it with stored breast milk.
  • To defrost, store overnight in the fridge to retain nutrients; use within 24 hours. In urgent cases, defrost in warm water.
  • Dispose of breast milk if it smells unpleasant.
  • Never heat breast milk in the microwave as it can heat unevenly and cause burns.
  • Always check food temperature before feeding.

Troubleshooting:

  • No suction: Check assembly, pump attachment placement, silicone tube insertion, suction level, and if the pump is switched on. Ensure no milk obstructs the silicone valve.
  • No milk pumped: Relax and try again. Reduce suction level if painful. Consult a doctor.
  • Pain during pumping: Reduce suction level. Consult a doctor.
  • Milk leaking: Remove and re-attach the silicone cushion, ensuring it's correctly positioned and secured. Lean your upper body slightly forward.
  • Altered/damaged components: Do not use corrosive cleaning agents. Temperature fluctuations can damage plastic. Stop using if parts are damaged.
  • Pump not switching on: Ensure correct assembly and connection to mains power.

Maintenance features:

  • Cleaning and sterilising:
    • Before first use and each use: Clean and sterilise all separate parts that come into direct contact with breast milk (except the motor unit and silicone tube) individually. Never sterilise the motor unit.
    • Unscrew pump attachment from bottle.
    • Remove silicone tube from pump attachment and motor unit.
    • Separate pump attachment from silicone membrane and silicone valve.
    • Sterilise the bottle and pump attachment (with silicone membrane and silicone valve) for five minutes in boiling water or using a steam steriliser.
    • Do not use the breast pump until all individual parts are completely dry.
    • Do not use antibacterial cleaning products, as they may damage the plastic.
    • Clean the motor unit carefully with a slightly damp cloth only. Do not use cleaning agents or solvents. Never submerge the motor unit in water.
    • If not using for a prolonged period, remove batteries.
  • General safety:
    • Keep packaging material and small parts away from children.
    • Ensure cables are not accessible to children.
    • Only use the included mains part plug.
    • Children must not play with the device; cleaning and user maintenance by children must be supervised.
    • Do not pull, twist, or bend the mains cable.
    • Ensure technical specifications of mains part plug match motor unit before first use.
    • Pull mains part plug from socket during operational faults, before cleaning, and when not in use.
    • Avoid contact with hot surfaces or sharp-edged objects.
    • Do not use additional parts not recommended by the manufacturer.
    • Store instructions safely and make them accessible to all users.
    • Include instructions when handing over to third parties.
    • The motor unit can be reused, but the pump set is for single-person use for hygiene reasons.
    • Use only on the female breast.
    • Never use when drowsy or asleep, or in the bath/shower.
    • Do not open or shake the breast pump in case of a fault; refer to troubleshooting or contact Customer Services.
    • The pump set requires no maintenance.
    • Repairs must only be carried out by Customer Services or authorised retailers. Opening or repairing the device yourself voids the warranty.
    • Never submerge the breast pump in water or touch with wet/damp hands.
    • If the breast pump falls into water, immediately remove the mains part plug.
    • Place on an even, stable surface during use.
    • Unwind the mains cable fully to avoid overheating.
    • Never place the breast pump or mains part plug in the microwave.
    • Remove mains part plug when not in use; do not leave plugged in.
    • Dispose of damaged mains cable.
    • Keep mains cable away from hot surfaces.
    • Not for use by individuals with reduced physical, sensory, or mental skills or lack of experience/knowledge unless supervised or instructed.
    • No calibration needed.
    • No modifications allowed.
  • Transport of feeding bottle:
    • Ensure screw plug is tightly screwed on when transporting breast milk.
    • Do not store filled bottle in suitcases, bags, etc. during transport.
  • Child safety with feeding bottle:
    • Always use with adult supervision.
    • Never use feeding teats as a soother.
    • Continuous/prolonged sucking of fluids causes tooth decay.
    • Always check food temperature before feeding.
    • Inspect feeding teat before each use; discard if damaged or weak.
    • Do not leave feeding teat in direct sunlight, heat, or sterilising solution for longer than recommended.
    • Do not tip teat in sweet substances.
    • Do not use solvents or abrasive cleaning agents.
    • Before first use, place milk-contacting parts in boiling water for 5 minutes for hygiene.
    • Clean before each use.
    • Liquid inside bottle must not exceed 50°C.
  • Battery handling:
    • If battery fluid contacts skin/eyes, rinse with water and seek medical assistance.
    • Choking hazard: Keep batteries out of reach of small children.
    • Observe polarity (+/-).
    • If a battery leaks, wear protective gloves and clean compartment with a dry cloth.
    • Protect batteries from excessive heat.
    • Risk of explosion: Never throw batteries into fire.
    • Do not charge or short-circuit batteries.
    • Remove batteries if device is not used for a long period.
    • Use identical or equivalent battery types only.
    • Replace all batteries at the same time.
    • Do not use rechargeable batteries.
    • Do not disassemble, split, or crush batteries.

Disposal:

  • Dispose of empty, flat batteries in designated collection boxes, recycling points, or electronics retailers. Batteries containing harmful substances (Pb, Cd, Hg) are marked accordingly.
  • For environmental reasons, do not dispose of the device in household waste. Dispose of at a suitable local collection or recycling point in accordance with EC Directive – WEEE.

Electromagnetic compatibility:

  • Complies with EU Medical Devices Directive 93/42/EEC and EN 60601-1-2.
  • Suitable for use in all environments listed, including domestic.
  • Use may be limited in the presence of electromagnetic disturbances, potentially causing error messages or display/device failure.
  • Avoid using directly next to or stacked on other devices, as this can lead to faulty operation. If necessary, monitor devices for proper function.
  • Using non-manufacturer specified accessories can increase electromagnetic emissions or decrease immunity, leading to faulty operation.
  • Failure to comply can impair device performance.

Beurer BY 40 Specifications

General IconGeneral
BrandBeurer
ModelBY 40
CategoryBreast Pump
LanguageEnglish

Related product manuals