Electromagnetic Compatibility Information
Table 1
For all me equipment and me systems
•ThedeviceneedsspecialprecautionsregardingEMCandneedstobeinstalledandputintoserviceaccordingtotheEMCinformation
provided.
•PortableandmobileRFcommunicationsequipmentcanaffectdevice.
•Theuseofaccessoriesorreplacementparts,otherthanthosesold/specifiedbythemanufacturermayresultinincreasedemissionsor
decreased immunity of the device.
•Thedeviceshouldnotbeusedadjacenttoorstackedwithotherequipmentandthatifadjacentorstackeduseisnecessary,thedevice
should be observed to verify normal operation in the configuration in which it will be used.
Guidance and manufacture’s declaration – electromagnetic emissions
EM34
is intended for use in the electromagnetic environment specified below. The customer or the user of
EM34
should assure that it is used
in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions CISPR 11 Group 1
EM34
uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B
EM34
is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2 Not applicable
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Not applicable