EasyManua.ls Logo

Bio-Tek FL600 - Scanning and Temperature Control

Bio-Tek FL600
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical Information
5-3
1.5 mm special probe for 384-well plates with
minimum crosstalk, reduced sensitivity on other
plates
1.0 mm special probe for high-density media.
Probe Selector Probe diameters can be different in the top read and
bottom read positions.
Selection of top or bottom read position is under software
control.
The height of the top reading probe may need to be set by
the user for different plate heights. See the section
entitled Adjusting Top Optical Probe Height in Chapter
3.
6FDQQLQJ0HFKDQLVP
Speed 65 mm/second maximum
Resolution 5 microns
Repeatability within ±25 microns
X Drive 10.0 mm leadscrew
Y Drives 10.0 mm leadscrew
There are two (Y axis) Probe scanning mechanisms fitted for top and bottom
read configurations
Probe Configurations Selection of top or bottom reader or optical density mode
under computer control