Proper device insertion, handling, and storage are critical to prevent instrument damage and ensure accurate results.
Swab a 10x10 cm area, ensuring the tip is moist. Reinsert swab into the body tube.
Push plunger to dispense reagents, then insert device into MVP ICON for reading within two minutes.
Readings of 2.5 Zones or less indicate a clean surface based on default thresholds.
Readings between 2.6 - 3.0 Zones suggest the surface is not adequately clean.
Readings greater than 3.0 Zones indicate a dirty surface requiring re-cleaning and re-testing.
Run ATP Positive Controls periodically. Expected result: 3.1-4.3 Zones.
Use unused activated device. Expected result: 1.0-2.5 Zones. Above 2.5 may indicate contamination.
Yellow band on swab is normal. Do not use if accidentally activated before sampling.
Store devices between 2°C-30°C (36°F–86°F), protect from sunlight, and reseal unused bags.
Warrants product free from defects in materials/workmanship when stored/used as intended.
Self-contained device for quality monitoring, detecting food and microbiological residues using ATP.
Contains one box of 100 surface sampling devices for use with MVP ICON/MVP systems.
Proper device insertion, handling, and storage are critical to prevent instrument damage and ensure accurate results.
Swab a 10x10 cm area, ensuring the tip is moist. Reinsert swab into the body tube.
Push plunger to dispense reagents, then insert device into MVP ICON for reading within two minutes.
Readings of 2.5 Zones or less indicate a clean surface based on default thresholds.
Readings between 2.6 - 3.0 Zones suggest the surface is not adequately clean.
Readings greater than 3.0 Zones indicate a dirty surface requiring re-cleaning and re-testing.
Run ATP Positive Controls periodically. Expected result: 3.1-4.3 Zones.
Use unused activated device. Expected result: 1.0-2.5 Zones. Above 2.5 may indicate contamination.
Yellow band on swab is normal. Do not use if accidentally activated before sampling.
Store devices between 2°C-30°C (36°F–86°F), protect from sunlight, and reseal unused bags.