50 MP System Hardware Guide
TSD105A Calibration
The TSD105A is easily calibrated using weights of known mass. Ideally, calibration should be performed with
weights that encompass the range of the forces expected during measurement and should cover at least 20% of the
full scale range of the transducer. When calibrating for maximum range on the force transducer, use weights that
correspond to 10% and 90% of the full scale range for best overall performance.
See DA100C Calibration options on page
45.
TSD105A Specifications
Rated Output: 1mV/V (normalized to 1V excitation)
Ranges: 50, 100, 200, 500, 1000 grams
Noise (rms): (Range/50)mg @ 10 volts excitation, 1Hz bandwidth
Nonlinearity: <±0.025% FSR
Hysteresis: <±0.05% FSR
Nonrepeatability: <±0.05% FSR
30 minute creep: <±0.05% FSR
Temperature Range: -10°C to 70°C
Thermal Zero Shift: <±0.03% FSR/°C
Thermal Range Shift: <0.03% Reading/°C
Maximum Excitation: 10 VDC
Mounting Rod: 9.5mm (dia) – variable orientation
Weight: 300g (with mounting rod)
Length: 19mm (wide), 25mm (thick), 190mm (long)
Cable Length: 3 meters
Interface: DA100C—see page