STATus:OPERation:ENABle
This command and its query can be used to set and read the value of the Operation Enable register.
This register is a mask for enabling specific bits from the Operation Event register to set the operation
summary bit (OPER) of the Status Byte register. The operation summary bit is the logical OR of all
enabled Operation Event register bits.
Command Syntax
STATus:OPERation:ENABle <NR1>
Parameters
0 to 65535
Default Value
0
Examples
STAT:OPER:ENAB 32 STAT:OPER:ENAB 1
Query Syntax
STATus:OPERation:ENABle?
Returned Parameters
<NR1> (register value)
Related Commands
STAT:OPER?
STATus:QUEStionable?
This query returns the value of the Questionable Event register. The Event register is a read-only
register that holds (latches) all events that pass into it. Reading the Questionable Event register clears
it.
Query Syntax
STATus:QUEStionable[:EVENt]?
Parameters
None
Examples
STAT:QUES:EVEN?
Returned Parameters
<NR1> (register value)
STATus:QUEStionable:CONDition?
This query returns the value of the Questionable Condition register. This is a read-only register that
holds the real-time (unlatched) questionable status of the electronic load.
Query Syntax
STATus:QUEStionable:CONDition?
Parameters
None
Examples
STAT:QUES:COND?
Returned Parameters
<NR1> (register value)