EasyManua.ls Logo

BK Precision 8601 - Page 46

BK Precision 8601
81 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
46
Command Syntax
[SOURce:]PROTection:CLEar
Parameters
None
Examples
:PROT:CLE
[SOURce:]INPut:TIMer
These commands enable or disable the electronic load on timer.
Command Syntax
[SOURce:]INPut:TIMer[:STATe] <bool>
Parameters
0 | 1 | OFF | ON
*RST Value
OFF
Examples
INP:TIM 1
Query Syntax
INPut:TIMer[:STATe]?
Returned Parameters
0 | 1
Related Commands
INP:TIM:DEL
[SOURce:]INPut:TIMer:DELay
This command specifies the timer setting.
Command Syntax
[SOURce:]INPut:TIMer <NRf+>
Parameters
1 to 60000s | MINimum | MAXimum | DEFault
Unit
seconds
*RST Value
10
Examples
INP:TIM:DEL 5
Query Syntax
[SOURce:]INPut:TIMer:DELay? [ MINimum | MAXimum |
DEFault ]
Returned Parameters
<NR3>
Related Commands
INP:TIM
[SOURce:]CURRent
This command sets the current that the load will regulate when operating in constant current mode.
Command Syntax
[SOURce:]CURRent[:LEVel][:IMMediate] <NRf+>
Parameters
0 through max rated current
Unit
A (amps)
*RST Value
MINimum
Examples
CURR 5 CURR:LEV 0.5
Query Syntax
[SOURce:]CURRent[:LEVel][:IMMediate]?
Returned Parameters
<NR3>
Related Commands
CURR:RANG