48
[SOURce:]CURRent:SLEW:POSitive
This command sets the slew rate of the current for positive transitions. MAXimum sets the slew to the
fastest possible rate. MINimum sets the slew to the slowest rate.
Command Syntax
[SOURce:]CURRent:SLEW:POSitive <NRf+>
Parameters
See specifications
Unit
A (amps per micro second)
Examples
CURR:SLEW:POS 0.0001
Query Syntax
[SOURce:]CURRent:SLEW:POSitive?
Returned Parameters
<NR3>
Related Commands
CURR:SLEW
[SOURce:]CURRent:SLEW:NEGative
This command sets the slew rate of the current for negative transitions. MAXimum sets the slew to the
fastest possible rate. MINimum sets the slew to the slowest rate.
Command Syntax
[SOURce:]CURRent:SLEW:NEGative <NRf+>
Parameters
see specifications
Unit
A (amps per micro second)
Examples
CURR:SLEW:NEG 0.0001
Query Syntax
[SOURce:]CURRent:SLEW:NEGative?
Returned Parameters
<NR3>
Related Commands
CURR:SLEW
[SOURce:]CURRent:PROTection:STATe
This command enables or disables the overcurrent protection feature.
Command Syntax
[SOURce:]CURRent:PROTection:STATe <Bool>
Parameters
0 | 1 | OFF | ON
*RST Value
OFF
Examples
CURR:PROT:STAT 1
Query Syntax
[SOURce:]CURRent:PROTection:STATe?
Returned Parameters
<NR3>
Related Commands
CURR:PROT
[SOURce:]CURRent:PROTection

Other manuals for BK Precision 8601