Unit Operation
Unit Operation
Unit Operation
Unit Operation
Unit Operation
18
20
17
19
21
Testing
1. Power On
"START/STOP"
The LCD screen will appear for one second as unit performs
a quick diagnosis. A long tone indicates device is ready for
testing.
Press and hold button until a beep sounds.
Note: Unit will not function if residual air from previous
testing is present in cuff. The LCD will flash " "
until pressure is stabilized.
2.
Initial pressure is first pumped to 190mmHg. If the current
user's systolic blood pressure is over 190mmHg, the unit
will automatically re-inflate to the proper shelf.
Pressurization
Note: Pressurization will gradually subside and ultimately
stop when cuff is not properly applied to the wrist.
If this occurs, pressSTART/STOP button to turn
the unit off.
3. Testing
After cuff inflation, air will slowly subside as indicated by the
corresponding cuff pressure value. A flashing " " will
appear simultaneously on screen signaling heart beat detection.
Note: Remain relaxed during testing. Avoid speaking or
moving body parts.
4. Result Display
Three short beeps sound when testing is complete.
The screen will display measurements for systolic and
diastolic blood pressure. An indicator representing
the current measurement will appear next to the
corresponding WHO Classification.
Irregular Heartbeat Indicator
If the monitor detects an irregular heart rhythm two or more
times during the measuring process, the Irregular Heartbeat
Symbol " "appears on screen along with measurement
results. Irregular heartbeat rhythm is defined as rhythm that
is either 25% slower or faster than the average rhythm detected
while measuring systolic blood pressure and diastolic blood
pressure. Consult your physician if the Irregular Heartbeat
Symbol " " frequently appears with your test results.
Indicator
Unit Operation
22
Note: Refer to Page 31~32 for detail WHO Blood Pressure
Classification Information.
mmHg
/
(
( )
)
(
( )
)
Do not stand while testing. Sit in a comfortable position and
place wrist level with heart.