15
ResMed S9 Series devices operate from a 24V direct current so the use of
a power converter or sine wave power inverter is required to complete the
connection between the device and the Battery Pack. DO NOT connect
the Battery Pack directly to any ResMed model device.
Battery Power Solutions’ 24V Power Converter was specifically designed
for use with ResMed S9 Series PAP devices.
Insert the 24V power converter into the DC out jack of the Battery
Pack. The 24V power converter must be connected to the Battery
Pack before it is connected to the ResMed S9 PAP device.
DUAL BATTERY PACK CONFIGURATION:
Insert the 24V power converter into the DC power pigtail cord.
Insert the DC power pigtail cord into the DC out jacks of
both Battery Packs.
Insert the 24V power converter into the DC input port of the S9 device.
Turn the PAP device on and it should begin functioning.
Powering Your PAP Device: ResMed S9 Devices
24V POWER CONVERTER
DC OUT JACK
PAP DC INPUT PORT
Note: ResMed’s S9 Series DC/DC Converter or a sine wave power inverter may also be
used to complete the connection between the ResMed S9 device and the Battery Pack.