Display Reference
39
Arithmetic overflow in
data calculation.
After a BP analysis error, the corrective action should
allow a return to normal operation for the next
measurement. If the problem persists, discontinue use
and return the monitor for servicing.
Temperature Errors (E41 to E55)
These error codes only apply to the BPM300. If a
temperature error occurs during any temperature
measurement, an error code will be displayed in the
Temperature display.
Continuous
temperature over
display limit (>43 °C)
Check monitor against reference
thermometer in water bath at 43 °C
to ensure monitor is working
properly.
Continuous
temperature under
display limit (<28 °C)
Check monitor against reference
thermometer in water bath at 28 °C
to ensure monitor is working
properly.
No temperature probe
connected
Attach temperature probe and repeat
measurement.
Timeout on single
temperature
measurement
Allow probe to cool to room
temperature for 34 minutes, press
Temperature Start, and then repeat
measurement.
Continuous
measurement
temperature circuitry
not operational
Internal malfunction. Contact BpTRU
Customer Support.