BpTRU Operator’s Manual
40
Single measurement
failed motion check
Repeat measurement and ensure that
probe is located correctly, patient
remains still during single
temperature measurement, and
patient’s mouth is closed. Use
Continuous mode if error is repeated.
Single measurement
failed curve check
Repeat measurement and ensure that
probe is located correctly, patient is
still during single temperature
measurement, and patient’s mouth is
closed. Use Continuous mode if
error is repeated.
Single measurement
over display limit
(>39 °C)
Repeat measurement. Use
Continuous mode if error is repeated.
Single measurement
under display limit
(<35 °C)
Repeat measurement. Use
Continuous mode if error is repeated.
Single measurement
start temperature out
of range (<10 °C or
>32 °C)
Allow probe to cool to room
temperature for 34 minutes, press
Temperature Start, and then repeat
measurement. Use Continuous mode
if error is repeated.
Single measurement
temperature circuitry
not operational
Internal malfunction. Contact BpTRU
Customer Support.
After a temperature error, the corrective action should
allow a return to normal operation for the next
measurement. If the problem persists, discontinue use
and contact your distributor or BpTRU Customer
Support.