Braun ExactFit 5 - Morning Hypertension Setting

Braun ExactFit 5
30 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
Morning average (for BP5900PH-EMEA only)
Press the average (3) button again into “morning
average
mode, the morning measurements
in memory within past 7 days will show on LCD.
If there is no data in memory within past 7
days, “
will show on LCD.
Evening average (for BP5900PH-EMEA only)
Press the average (3) button once again into evening
average
mode, the evening measurements in memory
within past 7 days will show on LCD.
If there is no data in memory within past 7 days,
will
show on LCD.
Erasing data
Press the memory
M
(2) button into memory mode.
Press and hold the mode
(5) button and then press the
“set
(4) button, all the stored data will be erased and
CLR will show on screen.
To check if the data was erased, press the memory
M
(2)
button. No data should show up on screen, but date, time
and M:00.
Morning hypertension setting (for BP5900PH-EMEA only)
Morning hypertension symbol will appear when morning average is above your
pre-congured systolic rate and/or pre-congured diastolic rate, and/or one or both of
these measurements is 20% or more above evening average.
** Your pre-congured systolic / diastolic should be decided between you and your
doctor or healthcare provider.
After setting date and time, press mode
(5) button into
“morning hypertension
setting mode “- - ” on systolic
position will ash for systolic setting.
Press set
(4) button, LCD will appear 135 mmHg for SYS
as the internal pre-set value, then press “Set
button
repeatedly to set his own value (range from 70~250 mmHg)
Press mode
(5) button again into morning hypertension
setting mode “ - - ” on diastolic position will ash for
diastolic setting.
Press “SET
(4) button, LCD will appear 85 mmHg for DIA
as the internal pre-set value, then user presses set
button
to set his own value (range for DIA is 40 ~ 115 mmHg).
User can press start/stop
(1) button to exit setting mode
and go back to standby at any step.