thermopress 400
Bitte diese Betriebs- und Wartungsanleitung vor dem Produkteinsatz ausführlich lesen!
Für Schäden, die durch Nichtbeachten dieser Betriebs- und Wartungsanleitung
entstehen lehnt der Hersteller jede Haftung ab!
Betriebs- und Wartungsanleitung Seite 1
Operation and maintenance manual page 13
Manuel d'utilisation et d'entretien page 25
Manuale di funzionamento e manutenzione pagina 37
Manual de funcionamiento y mantenimiento paginá 49
Manual de funcionare i întreinere pagina 61
       73
DE
EN
FR
IT
RO
ES
RU
009183EX-20181220

Table of Contents

Questions and Answers

  • A
    Andres JenkinsAug 13, 2025
    Why is the injection performance decreasing in my bredent thermopress 400?
    • B
      Brittany AcostaAug 13, 2025
      Decreased injection performance in your bredent Dental equipment can occur if the heating chamber or the front side of the pressure piston is contaminated, so clean them. Also, ensure you're using sufficient thermopaste for aluminium cartridges, as too little can increase cartridge friction. Finally, verify and correct your program settings to match the parameters for injection molding.
  • K
    kevinchandlerAug 17, 2025
    Why is my bredent Dental equipment not reaching the desired temperature?
    • M
      Michael WillisAug 17, 2025
      If your bredent Dental equipment isn't reaching the desired temperature, first compare the mains voltage with the unit's technical data. Check the heating elements for defects. If the temperature probe is inaccurate, calibrate it using a digital thermometer and temperature sensor (accessories 99300364 and 99300366). Regular checks of the target/actual temperature will ensure flawless injection molding. Also, check and adjust the parameters for the thermoplastic material.
  • J
    Jeffrey MartinAug 20, 2025
    What to do if bredent thermopress 400 shows 'Motor temperature too high!'?
    • J
      Jenna MercerAug 20, 2025
      If you receive a 'Motor temperature too high!' message on your bredent Dental equipment, the thermal protection switch has been activated, so the unit can no longer be used. Make sure the unit is in a well-ventilated area. You can also abort the program using the CE key and allow the injection moulding unit to cool down while still switched on (ventilation mode).
  • N
    Norma FrederickAug 25, 2025
    How to fix a blank display on my bredent Dental equipment?
    • C
      Catherine MooreAug 25, 2025
      If the menu selection isn't showing on the display of your bredent Dental equipment after pressing the main switch, check the fuses for the power supply and the cable connections to the unit. Ensure the mains voltage corresponds to the value on the type-plate and that the socket is live. You may also need to change the unit fuse (10 [A] microfuse).
  • J
    Jamie RomeroAug 28, 2025
    What to do if the injection process won't start on bredent thermopress 400 Dental equipment?
    • S
      Sydney VasquezAug 28, 2025
      If the injection process won't start on your bredent Dental equipment, verify that the position of the heating cylinder and its filling matches what's shown on the display. Make sure the heating cylinder was brought to the final position when fixing the flask. Use the selection lever to move the heating cylinder to the desired final position (1 or 2) until it stops. Ensure the flask chamber lid (protective lid) is correctly closed and that the right rear contact switch is functioning correctly.
  • H
    hansenmelissaSep 1, 2025
    Why is there aluminium residue in my bredent thermopress 400 Dental equipment's heating cylinder?
    • B
      Brandon BrandtSep 1, 2025
      If you find a large amount of aluminium residue in the heating cylinder of your bredent Dental equipment, ensure you are applying sufficient thermopaste to the aluminium cartridge. Also, check for increased abrasion of the aluminium cartridges. If necessary, have the heating block replaced.
  • X
    xsmithSep 4, 2025
    Why does my bredent thermopress 400 Dental equipment display incomprehensible messages?
    • S
      Sharon StevensSep 4, 2025
      If you are seeing incomprehensible messages on the display screen of your bredent Dental equipment, make sure the unit is set up in an adequately ventilated area. You can also try switching the unit off, waiting approximately 2 minutes until a quiet "click" is heard, and then restarting it.
  • L
    Lance Shepherd MDSep 8, 2025
    Why does my bredent Dental equipment say "Cooling down"?
    • M
      Michael SmithSep 8, 2025
      If your bredent Dental equipment displays "Cooling down", wait until the unit cools down and the temperature is reached. This occurs when the target temperature programmed for inserting the filled aluminium cartridges is below that of the heating cylinder.