OPERATING INSTRUCTIONS
Page 1 of 2
WARNING
The Polar Care Cube can be cold enough to cause serious injury, including
full skin necrosis. Follow these Operating Instructions, and carefully read the
Product Insert (see pouch on side of unit) and the Cold Therapy Pad Fitting
Instructions (provided with each Cold Therapy Pad) prior to use.
1 Discuss Treatment with your Licensed Health Care Practitioner
Provide a complete medical history including any reactions to cold. Certain medical conditions make cold-induced injury more likely. Ask your
practitioner about potential adverse reactions and cold-induced injuries.
2 Use Only as Prescribed
Use only according to your practitioner’s instructions regarding the frequency and duration of cold application and length of breaks between uses,
how and when to inspect the skin, and total length of treatment. Do not use this device if you did not receive or do not understand the instructions.
Unless your practitioner provides dierent instructions, to take a break between uses simply disconnect the power from the unit or remove the pad
from your body for a minimum of 30 minutes. Federal law restricts this device to sale by or on the order of a licensed health care practitioner.
3 Apply Insulation Barrier & Cold Therapy Pad
Do not let any part of the pad touch skin. Always use an insulation barrier (such as Breg Polar Dressing, Webril™, Kerlix™, cast padding or elastic
bandage) between the Cold Therapy Pad and skin. If a sterile dressing has been applied to the treatment site that does not completely cover the
skin under the pad, use an additional insulation barrier. Use only with the Breg Cold Therapy Pads. Other pads may be colder, increasing the risk
of serious cold-induced injury, including full thickness necrosis.
4 Check for Moisture
Check for moisture on the barrier between your skin and the cold pad. Discontinue use if the barrier is moist. Change to a dry skin barrier before
resuming use.
5 Inspect Skin Regularly
Inspect the skin under the Cold Therapy Pad (by lifting the edge) as prescribed, typically every 1 to 2 hours. Ask your practitioner to instruct you
on how to inspect the skin area which is being treated by the device. Do not use the Polar Care unit if dressing, wrapping, bracing, or casting
over the Cold Therapy Pad prevents skin checks. Stop using and contact your practitioner immediately if you experience any adverse reactions,
such as: increased pain, burning, increased swelling, itching, blisters, increased redness, discoloration, welts, other changes in skin appearance, or
any other reaction identied by your practitioner.
Awake/
Asleep
Asleep
Asleep
Asleep
Awake
Awake
Awake
Day:
Day:
Day:
Day:
Day:
Day:
Through
Through
Through
Treatment
Period
Frequency/Duration
Inspect Skin
Every:
Cold Therapy Protocol*
* To be completed by a licensed Health Care Professional
Set Up
Coupler Connections
Connect
FIRST ADD WATER TO
FILL LINE, THEN ICE
Replace ice when existing
ice is almost melted
Operation
End Session
To stop the pump, disconnect electrical power.
Remove pad.
See Product Insert for
electrical precautions.
Check Skin Frequently
Over
POLAR CARE
®
CUBE
TM
Question and Answer IconNeed help?

Do you have a question about the Breg polar care cube and is the answer not in the manual?

Breg polar care cube Specifications

General IconGeneral
TypeCold Therapy System
CategoryMedical Equipment
Cooling Reservoir Capacity8.5 Quarts (8 Liters)
Capacity8.5 Quarts (8 Liters)
Weight (empty)4.6 lbs (2.1 kg)
Cooling Duration6-8 hours
CompatibilityBreg Cold Therapy Pads
Dimensions9 in x 9 in x 9 in (23 cm x 23 cm x 23 cm)
Power SourceAC power

Summary

OPERATING INSTRUCTIONS

Discuss Treatment with your Licensed Health Care Practitioner

Discuss medical history and potential cold reactions with your practitioner for safe usage.

Use Only as Prescribed

Follow practitioner's instructions for frequency, duration, and skin inspection. Federal law restricts sale.

Apply Insulation Barrier & Cold Therapy Pad

Use an insulation barrier between the pad and skin to prevent cold injury. Use only Breg pads.

Check for Moisture

Monitor moisture on the barrier; change to a dry barrier if moist before resuming use.

Inspect Skin Regularly

Inspect skin every 1-2 hours. Stop use and contact practitioner if adverse reactions occur.

Tips and Trouble Shooting

Usage Tips

Use cubed ice, drain pad between uses, disconnect hose coupling to remove pad, do not run without water.

Pump not running, water not flowing to pad, or pad not cold

Check ice size, flow stabilization, power, water level, hose kinks, connections, and clean filter.

Condensation

Minimize air exposure with wrap material or use a sterile dressing with waterproof barrier.

Unit is leaking

Check connectors, apply lubricant to O-rings, or contact customer care if leaks persist.

Cleaning

Empty and dry unit after use. Clean pump and tubes with mild detergent and warm water if needed.