The 20g Bent Spray supplied with each Unit allows Open Spraying in any position through 360 degrees and
eliminates the need to tip the Unit.
This Cryosurgical Unit is designed only for use with other Brymill manufactured products.
Your unit is supplied with 4 different sizes of Open Spray Apertures, a 20G X 1.0in Straight Spray, and a 20g Bent Spray.
The full range of Open Sprays and Closed Probes can be found on our website. Your selection of Open Spray or Contact
Probe will depend upon the size and type of lesion being treated.
Once the bottle is filled with liquid nitrogen, attach the appropriate Spray Tip or Probe to the permanently affixed Knurled
Nut by turning the spray tip or probe until the threads are fully engaged with finger tight firmness. Depress the finger
trigger to dispense and control the flow of liquid nitrogen.
When you have completed the treatment of a patient, set the Cryosurgical Unit gently on a table. The bottom of the unit
may be damaged if it is dropped or repeatedly brought in contact with a hard surface.
At the conclusion of an office day, the Cryosurgical Unit should be emptied and dry. To ensure that all liquids are removed
from the bottle, the components should air-dry separately, with the bottle and cap inverted. Any foreign materials detected
inside the bottle may be removed by wiping or rinsing with 70% IPA alcohol. Do not attempt to use the instrument until the
bottle is completely dry. Bottle should be stored in a CLOSED position (with the top on).
10. CLEANING/DISINFECTION INSTRUCTIONS
After each patient use follow the cleaning/disinfecting instructions provided below for the Cry-Ac
®
, Cry-Ac-3
®
and Cry-
Baby
®
.
1.Prepare an enzymatic, neutral pH cleaner solution according to the Manufacturer’s instructions.
2. Immerse a clean, lint-free wipe into cleaning solution and thoroughly wring. Thoroughly wipe bottle and top
assembly surfaces in a circular motion. Discard the wipe.
3. Using a soft, nylon brush, scrub difficult-to-access areas such as crevices or textured surfaces.
4. Immerse a fresh, clean, lint-free wipe under warm flowing, utility water and thoroughly wring excess water.
Thoroughly wipe the front panel surface and crevices for at least 30 seconds. Discard the wipe.
5. Rinse a fresh, clean, lint-free wipe under warm flowing, utility water and thoroughly wring excess
water. Wipe for at least 30 seconds. Discard the wipe.
6. Inspect the device for visible soil or debris. If visible soil remains, repeat cleaning steps 2-5 until device is
visually clean.
7. Using disinfection wipe thoroughly wipe bottle and top assembly surfaces and discard the wipe. Allow the wiped
and wetted device to stand for a minimum of the disinfectant product manufacturers labeled
contact time. Re-wet the wipes as necessary to ensure that all surfaces remain wet the entire time.
8. Use a 70% IPA wipe, or a low-lint wipe, saturated with 70% IPA, to thoroughly wipe the device.
9. Allow the device to air dry. Once dry, visually inspect the device for any residual soil.
10. Repeat cleaning/disinfection instructions if visible soil is present.
11. Place the cleaned/disinfected device in an appropriate dry storage area associated sprays and probes are
cleaned at the end of the Clinic Day.
When used with open sprays, the Cry-Ac
®
, Cry-Ac-3
®
or Cry-Baby
®
and the associated accessories do not come into
direct contact with the patient; therefore, the risk of infection is low and the unit and accessories may be cleaned. If the
products are intended to be used in the sterile field, they may be autoclaved per the recommendations below.
Contact Probes are directly in contact with the patient, please refer to the Instructions for Use supplied with each Contact
Probe or go to the following link for specific instructions on cleaning and sterilization:
http://www.brymill.com/docs/default-source/PDFs/contact-probes-instruction-for-use.pdf?sfvrsn=2