7 – 52
7 Maintenance
KjelDigester K-446 / K-449 Operation Manual
7. If the fuse brakes repeatedly contact the customer support.
∙ 8.1 Customer service, (8–55).
7. 5 Rechargeable battery (K-449 only), replacement
aRequired tools:
• TORX screw driver T10
bPreliminary tasks
1. Remove the drip tray and store it on the drip tray holder on the left side of the device.
2. Remove the suction module (7–54).
cTasks
1. Make sure the lift is in the cooling position.
WARNING
Risk of electric shock with connected power supply cable!
Serious injuries or death can result.
1. Switch off the device.
2. Disconnect the power supply cable from the device.
2. Switch off the device.
3. Disconnect the power supply cable from the device.
4. Remove the 4TORX screwsa of the lift cover.
Required tools
TORX screw driver T10
5. Slide out the lift coverb.

Table of Contents