Gas Analysis
Installation and Operation Instructions
Original instructions
Sample gas cooler
TC-MIDI (+)
BE440025
10/2018
Bühler Technologies GmbH, Harkortstr. 29, D-40880 Ratingen
Tel. +49 (0) 21 02 / 49 89-0, Fax: +49 (0) 21 02 / 49 89-20
E-Mail: analyse@buehler-technologies.com
Internet: www.buehler-technologies.com

Table of Contents

Questions and Answers

  • C
    Christopher GarrettAug 11, 2025
    Why is my Buhler Freezer showing excess temperature?
    • O
      oweberAug 11, 2025
      The Buhler Freezer may show excess temperature because the operating point has not yet been reached (wait max. 20 min), the cooling outlet is too long despite the cooler running (be sure the vents are not covered to prevent heat buildup), the flow rate, dew point, or gas temperature is too high (maintain limits and/or install pre-separator), or the installed fan has stopped (check and replace if necessary).
  • A
    Anthony RothAug 19, 2025
    What to do if there is reduced gas flow rate in Buhler TC-MIDI Freezer?
    • E
      ericksonpatriciaAug 19, 2025
      If you experience a reduced gas flow rate in your Buhler Freezer, it could be due to a clogged gas circuit (uninstall and clean the heat exchanger, and if necessary, replace the filter element) or a condensate outlet iced over (send in unit).
  • W
    wilsonmaryAug 26, 2025
    How to fix condensate inside the gas outlet of a Buhler TC-MIDI?
    • C
      Chase ArmstrongAug 26, 2025
      If you find condensate inside the gas outlet of your Buhler Freezer, the condensate trap may be full (empty the condensate trap), the valve inside the automatic condensate drain may be stuck (flush in both directions), or the cooler may be overloaded (maintain limits).
  • J
    Jennifer WolfeSep 3, 2025
    Why Buhler Freezer has insufficient temperature?
    • K
      Keith PetersonSep 3, 2025
      Insufficient temperature in the Buhler Freezer may be caused by a faulty control (send in cooler).