Measure what you see.
BYK-Gardner GmbH · Lausitzer Strasse 8 · 82538 Geretsried · Germany · Tel. +49-8171-3493-0 · Fax +49-8171-3493-140 · www.byk.com/instruments
Please note the detailed operating instructions in English on CD.
ENGLISH
haze-gard plus Operation
On the left Side:
• Main Switch on/off
• Voltage Selector
• Fuse
Connection for:
• Power Cable
• Foot Switch
• PC
How to Calibrate?
Calibration is necessary at the first
startup and at a regular interval
(two months recommended):
• Press set up button to call up
the calibration routine
• Cover haze-port with dark
standard and press operate
• Remove cover from haze-port
and start the 100%
measuremen
• Place the clarity standard at the
clarity port and press operate
operate button:
Measure
set up button:
SET-UP MENU
reference button:
Measure reference in Continuous
mode
Cursor keys:
Used in SET-UP mode to select
individual menu items and set
values.
In any measuring mode, these keys
are used to switch statistics
functions on/off, and to delete
values from statistics memory
SET UP MENUE
¢
Calibrate
£
Print Statistic
£
Print Online: off
£
Cont. Mode: off
£
Language USA
Escape: SET UP