13. Using the device in environmental conditions different than those indicated in this manual may harm seriously the safety
and the technical characteristics of the same;
14. The materials used to contain the drugs are made with highly stable thermoplastic polymers that are resistant against
chemicals. Such materials were tested with commonly used drugs (Salbutamol, Beclametasone dipropionate, Acetylcysteine,
Budesonide, Ambroxol) and no interaction phenomenon was observed. Interactions cannot however be excluded given the
variety and the continuous evolution of the drugs that are used. Remember to:
To consume the drugs as quickly as possible after opening its package;
To avoid keeping the drug in the tray-like container for too long and to clean it immediately after every application;
If the tray-like container presents any abnormal situation (such as softening or cracks), do not introduce any solution
and do not proceed with the inhalation. Contact the technical service and describe the methods and type of drugs
used.
15. Remember to:
Only use this device with medicines prescribed by your doctor;
Carry out the treatment only using the accessory indicated by the doctor according to the pathology.
TECHNICAL CHARACTERISTICS
Model
CLINEB PRO
Typology (MDD 93/42/EEC)
Class IIa Medical device
Power Feeding
230V ~ / 50 Hz
Power Consumption
170VA
Fuse
F 1 x 1.6A L 250V
Max Pressure
300 kPa (3.0 Bar)
Max Air Flow
16 l/min
Operating Pressure
95 kPa (0.95 Bar)
Operating Air Flow
8.0 l/min a 95kPa
NEBULIZZAZIONE (**)
Pos. A
(closed)
Pos. B
Pos. C
Pos. D
0,40 ml
0,60 ml
0,70 ml
0,80 ml
MMAD (µm) *
Mass Median Aerodynamic Diameter
Pos. A
(closed)
Pos. B
Pos. C
Pos. D
3.32
4.07
4.23
4.18
GSD (*)
Geometric Standard Deviation
Pos. A
(closed)
Pos. B
Pos. C
Pos. D
4.12
2.64
2.74
2.49
Output Rate (ml/min) *
Pos. A
(chiuso)
Pos. B
Pos. C
Pos. D
0.077
0.105
0.124
0.138
Output (ml) *
Pos. A
(closed)
Pos. B
Pos. C
Pos. D
0.189
0.237
0.247
0.260
Weight
2.20 Kg
Size
230 x 250 (H) x 190mm
Noise Level (measured as specifications of EN 13544-1)
Approx. 60dB (A)
Duty Cycle
Non-Stop Operated
Min Capacity Nebulizer
2ml
Max Capacity Nebulizer
8ml
Working Condition
Room temperature: 5 ÷ 40 °C
Room humidity percentage: 10 ÷ 93 % RH
Atmospheric pressure: 700 ÷ 1060 hPa
Conservation condition and Transport
Room temperature: - 25 ÷70 °C
Room humidity percentage: 0 ÷ 93% RH
Atmospheric pressure: 500 ÷ 1060 hPa
(*) Data determined with Cascade Impactor 290 series, compliant with EN 13544-1, by spraying 2 ml of NaF 1.0%.
(**) Data from free nebulisation of 2 ml of NaCl 0.9% (average nebulisation value per minute).