Memory: After pressing the
C
button a second time, the Falcon will show the last saved value or if
the PeakMin-Detect is on, the peak and minimum pressure. In this case the two values will be shown by
alternating the display every 3 seconds (The lower value is the minimum and the higher the peek pressure).
Setup: Pressing the
C
button a third time will allow you to change the settings of the Falcon.
Changing the Engineering Units: kPaa, kPag, psia, psig
Press the
C
button until Setup is displayed.
Press the
M
button to display Units, and the current unit will be displayed.
Press the
9
button, and the units will toggle between psia, psig, kPa and kPag.
Press the
M
button to save the desired unit.
Reset the Zero Offset: This useful if the Falcon has been accidentally zeroed under pressure, or
if you want to see how much the deadweight has shifted from the factory zero.
Press the
C
button until Setup is displayed.
Press the
M
button to display Units.
Press the
C
button to display Reset.
Press the
M
button to reset the gauge back to factory zero.
PeakMin-Detect: This feature changes how the memory mode works. If the PeakMin-Detect is
ON, the deadweight will display the last peak and minimum pressure sampled (instead of showing
the last saved pressure). When memory is displayed the peak and minimum pressure will be shown,
alternating every 3 seconds. By pressing the
M
key when the deadweight is in sampling mode,
and it will clear the peak and minimum pressure, so new pressures can be recorded.
Press the
C
button until Setup is displayed.
Press the
M
button to display Units.
Press the
C
button to display Reset.
Press the
C
button to display Pon or Poff.
Press the
9
key to toggle between Pon or Poff.
Sleep-Mode: Changes how the Falcon automatically turns off. Normally, the sleep-mode is set to
toff. In toff, the deadweight will stay on as long as there is more than 2 pounds of pressure. As
soon as there is less than 2 pounds of pressure, it will shut off after 5 minutes of inactivity. If the
Falcon has been set to ton, it will shut off in 5 minutes, after you stop pressing any of the buttons.
Press the
C
button until Setup is displayed.
Press the
M
button to display Units.
Press the
C
button 3 times, until toff appears.
Press the
9
key to toggle between toff and ton modes.
Press the
M
key to save the desired mode.