Model MCH-1000
Operator/Service Manual
Cardiopulmonary Bypass Temperature Controller
Publication date: June 15, 2022
LBL-01 V5
P/N 1000-05-1178

Table of Contents

Questions and Answers

  • J
    Jennifer ClarkSep 23, 2025
    What does a Line Isolation Monitor (LIM) alarm or error mean on CardioQuip Medical Equipment?
    • C
      claytonstewartSep 23, 2025
      If you receive a Line Isolation Monitor (LIM) alarm or error on your CardioQuip Medical Equipment, contact the LIM vendor if the LIM is not rated for the power requirements and/or power factor of the MCH. If the LIM is properly rated, contact CardioQuip Service, as the issue may be due to excessive leakage current from the MCH.
  • J
    Jose CraneAug 23, 2025
    What to do if there is no circulation on main water connections in CardioQuip Temperature Controller?
    • D
      Daniel RiveraAug 25, 2025
      If there is no circulation on the main water connections of the CardioQuip Temperature Controller, first ensure the system is in RUN mode to activate the pump. Check that the Internal Recirculation Valve is set to the NORMAL position. Also, inspect the water tank for blockages by cleaning the mesh strainers at the bottom of the tank with a nylon brush. Examine the external tubing for any closed valves, kinks, bends, or other obstructions. Other possible causes include a pump wiring fault, or a faulty pump or pump circuitry.
  • G
    guerrerovincentAug 25, 2025
    What to do if CardioQuip Temperature Controller system water flow is inadequate?
    • E
      Edward StanleyAug 26, 2025
      If the CardioQuip Temperature Controller system water flow is inadequate, place the system in PRIMING mode according to the operating instructions. Additionally, clean the mesh strainers at the bottom of the water tank with a nylon brush to remove any blockages. A blockage in the plumbing circuit may also be the cause, so have qualified service personnel inspect the internal tubing for kinks, bends, or other blockages. Another possible cause is a faulty pump or pump circuitry.
  • L
    Lisa OrtizAug 27, 2025
    Why is my CardioQuip Temperature Controller system not cooling?
    • R
      rodriguezdeborahAug 27, 2025
      If the CardioQuip Temperature Controller system will not cool, ensure there is an adequate supply of ice in the water tank, using ice made from sterile or 0.22-micron filtered water. Also, check if the water flow is inadequate. Finally, check if there is no water flow at the main water output.
  • K
    kimberlymorseAug 27, 2025
    What causes a CardioQuip Temperature Controller system to not heat?
    • P
      Paul JohnsonAug 27, 2025
      If the CardioQuip Temperature Controller system will not heat, have qualified service personnel inspect the internal tubing for kinks, bends, or other blockages. Additionally, have them check for an open in the Heater input wiring. Another possible cause is a faulty heater or heater circuitry.
  • J
    Jimmy Craig PhDSep 15, 2025
    Why doesn't water flow through the cold water tank in CardioQuip MCH-1000 Medical Equipment?
    • S
      sawyerjosephSep 15, 2025
      If water does not flow through the cold water tank of your CardioQuip Medical Equipment, have qualified service personnel inspect the internal tubing for kinks, bends, or other blockage, as this could indicate a blockage in the plumbing circuit. Additionally, they should check for an open in the valve input wiring to ensure internal wiring continuity and inspect and lubricate the valve solenoid to address potential valve solenoid seizure. For a faulty valve or valve circuitry, contact CardioQuip Service.
  • H
    hughessaraSep 12, 2025
    What to do if CardioQuip Medical Equipment's touchscreen display is off but the main power switch lights up?
    • J
      Jon ToddSep 13, 2025
      If the main power switch lights up, but the touchscreen display is off and the CardioQuip Medical Equipment system will not operate, have qualified service personnel inspect the internal wiring for disconnected plugs or loose wire connections, as this could be due to a disconnected DC Power Supply, or other loose internal wiring. If the wiring is intact, the issue may be a faulty DC Power Supply, in which case you should contact CardioQuip Service.
  • Z
    zthorntonSep 14, 2025
    What causes CardioQuip MCH-1000 Medical Equipment to go into an overtemperature condition?
    • M
      Michael OlsonSep 14, 2025
      If your CardioQuip Medical Equipment system goes into an overtemperature condition, have qualified service personnel inspect the internal tubing for kinks, bends, or other blockage. It could also be caused by faulty temperature control circuitry or a faulty heater or heater circuitry, in which case you should contact CardioQuip Service.
  • D
    donna45Sep 5, 2025
    Why are the hoses on my CardioQuip Temperature Controller cloudy or discolored?
    • E
      espinozaalexandraSep 5, 2025
      If the hoses of the CardioQuip Temperature Controller are cloudy or discolored, immediately remove the device from service and contact CardioQuip Customer Service regarding servicing assistance, as the device may be contaminated or the exterior tubing may have undergone physical changes.
  • R
    Ryan HobbsSep 2, 2025
    Why won't my CardioQuip Temperature Controller system operate, even though the touchscreen display is on?
    • J
      Jeffrey HarrisonSep 3, 2025
      If the CardioQuip Temperature Controller system will not operate, but the touchscreen display is on, have qualified service personnel check for an open in the DC Power Supply input wiring.

Summary

Warnings and Precautions

Specifications

Storage and Handling

Operation Overview

CONTROL SCREENS

System Test

Standby Mode

Priming Mode

Settings Screen

Defrost and Run Modes

ALARMS

Priority Alarms

Covers low water level and low ice alarms, detailing priority and silencing.

Overtemp Alarm

System Failure Alarm

Setting Temperature: 3 Ways

CONNECTORS, VALVES, & HOSES

Main Hose Connectors

Explains the large dripless connectors for arterial/cardioplegia heat exchangers.

Cardioplegia Hose Connectors

Internal Recirculation Valve

Blanket Hose Connectors

Drains

Blanket Orientation

Refrigeration Module

Refrigeration Module Maintenance

MCH OPERATION

Initial Operation Check

Control Module Check

Basic Function Test

System Tests

Temperature Control System Test

Explains how to test the temperature control system's accuracy and response.

Alarm Tests

Guides the user through testing various alarm functions of the MCH.

Operation Termination and Checks

Normal Operation

Overtemp Condition

MAINTENANCE

Water Quality and Cleaning

Covers water quality requirements and approved cleaners/disinfectants.

After Every Use Maintenance

Weekly Maintenance Procedure

Quarterly Maintenance Procedure

Annual Maintenance and Disposal

TROUBLESHOOTING

Common System Issues

Addresses problems like failure to turn on, unresponsive display, and circulation issues.

Troubleshooting: Flow Issues

Troubleshooting: Performance Issues

Wiring Schematics

MCH-1000(i) Schematic

Displays the detailed wiring schematic for the MCH-1000(i) model.

MCH-1000(m) Schematic

CardioQuip MCH-1000 Specifications