78 6H4866-02
[3] Confirm that the Left Limit Sensor functions by performing the following test:
(a) Enter the Diagnostic screen in the Kodak QC PoC 140.
(b) In the Linear Motor section, click the right or left arrow according to the limit switch you wish to test.
(c) The Carrier begins to move, and should enter the sensor area causing the indicator light to illuminate.
Linear Motor Section of Diagnostics Screen
[4] Assemble the scanner cover.
Indicator
Light
Indicator
Light