Instructions for Use
DualFlex
Hybrid Alternating Pressure Mattress Replacement System
for Dual Stage Nursing
U-2S2602-30631-G V1.2
DualFlex Hybrid Alternating Pressure Mattress Replacement System for Dual Stage Nursing

Questions and Answers

  • J
    Jose MossAug 12, 2025
    What to do if the Carilex DualFlex shows a low pressure indicator?
    • M
      mariajohnsonAug 12, 2025
      If the Carilex Medical Equipment shows a low-pressure indicator, first ensure that the air filter isn't blocked by cleaning it. If the issue persists, there may be leaks in the compressed air system or a defective air pump. In this case, move the patient to another bed while you contact your medical distributor's technical service for assistance.
  • T
    Timothy YoungAug 19, 2025
    Why Carilex DualFlex power unit is not working?
    • S
      Steven McmillanAug 19, 2025
      If your Carilex Medical Equipment power unit isn't working and there are no indications on the control device, start by ensuring the power unit is correctly plugged into a functioning wall outlet. If it still doesn't work, the power cord or fuse might be defective. Contact your medical distributor's technical service for assistance.
  • S
    Sean BooneAug 27, 2025
    How to troubleshoot Carilex DualFlex Medical Equipment when the power unit works but the mattress doesn't inflate?
    • K
      Karen ThompsonAug 27, 2025
      If the power unit of your Carilex Medical Equipment is working but the mattress isn't inflating, it could be due to damaged air cells or compressed air hoses. Begin by opening and removing the coverlet to inspect the compressed air hoses at their connection to the air cells, checking their positioning. Also, check the tightness of the compressed air hoses at the power unit. If problems persist, contact your medical distributor's technical service for assistance.
  • T
    Thomas WangSep 2, 2025
    What to do if there is no alternating pressure regulation in Carilex Medical Equipment?
    • I
      icainSep 2, 2025
      If your Carilex Medical Equipment has no alternating pressure regulation, this could be caused by a failure of the power unit, or bending and pressing points in the compressed-air hoses. Open and remove the coverlet, check the correct positioning of the compressed air hoses. For power unit failures, contact your medical distributor's technical service.