Weight 11.6 oz (329 g) with cuff,
without batteries
Cuff size
8.7 - 16.5 (22 - 42 cm)
Operating
Range
50° F - 104° F (10° C - 40° C),
10 - 85% relative humidity
(non-condensing)
Storage Range -4° F - 131° F (-20° C - 55° C),
10 - 90% relative humidity
Power supply 4 x 1.5 V
AAA batteries
Battery life Approx. 500 measurements
Expected
Service life
500hours
Classification Internal supply, IP21, no AP or APG,
continous operation, application part type BF
Technical information is subject to change without notification
to allow for updates. The serial number is located on the device
or in the battery compartment.
This device complies with European Standard EN 60601-
1-2 (in accordance with CISPR 11, IEC 61000-4-2, IEC
61000-4-3, IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-6,
IEC 61000-4-8, IEC 61000-4-11) and is subject to particular
precautions with regard to electromagnetic compatibility.
The device is suitable for use in all environments listed in
these instructions for use, including domestic environments.
Please note that portable and mobile HF communication
systems may interfere with this device.
The device complies with the EU Medical Devices
Directive 93/42/EEC, the German Medical Devices Act
(Medizinproduktgesetz) and the standards EN 1060-3 (Non-
invasive sphygmomanometers – Part 3: Supplementary
requirements for electro-mechanical blood pressure
measuring systems) and IEC 80601-2-30 (Medical electrical
equipment – Part 2 – 30: Particular requirements for the basic
safety and essential performance of automated non-invasive
sphygmomanometers).
The accuracy of this blood pressure monitor has been
carefully checked and developed with regard to a long useful
life. If the device is used for commercial medical purposes, it
must be regularly tested for accuracy by appropriate means.
Precise instructions for checking accuracy may be requested
from the service address.
12. Power Supply Unit
Input
100 240 V, 50 60 Hz, 0.5 A max.
Output
6 V DC, 600 mA, only in connection with
CaringMill blood pressure monitor.
Protection
This device is double insulated and protected
against short circuit and overload by a primary
thermal fuse
Make sure to take the batteries out of the
compartment before using the mains adapter.
Polarity of the the DC voltage connection
Double insulated / equipment class 2