30
4.1.3 Capacity & Units (Menu Code 8130)
(Calibration MENU -> 1. Calibration -> 3. Capacity & Units)
Set scale’s Weighing unit, capacity, Interval, Cal Unit.
Caution: Span calibration must take place after “Capacity & Units” setting.
Do not change setting after Span calibration.
Option Setting Value
1 Weighing Unit Setting Scale Unit
0 : kg
1 : lb
2 : g
2 Capacity Setting Scale Max Capacity
0 : 6 kg / 15 lb
1 : 15 kg / 30 lb
2 : 30 kg / 60 lb
3 Interval Setting Usage of Multi-interval
0 : Single Interval
1 : Dual Interval
4 Cal Unit Setting Calibration Weighing unit
0 : kg
1 : lb
* This setting uses in Span Calibration,
Percent Calibration, Linearity Adjust.
8130
CAL
ModE
Weighing Unit:[0] 0)kg 1)lb 2)g
Capacity :[1] 1)15kg
Interval :[1] 0)Single 1)Dual
Cal Unit :[0] 0)kg 1)lb